FDA Adverse Event Injury Summary report: N

BLUNT TIP SCREW, ÿ 4X48MM

MDR report key: 19351143 · Received May 20, 2024

Report

Report Number
0009613350-2024-00184
Event Type
Injury
Date Received
May 20, 2024
Date of Event
April 22, 2024
Report Date
June 20, 2024
Manufacturer
ZIMMER GMBH
Product Code
HSB
UDI-DI
00889024505483
PMA / PMN Number
K200814
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(4). D4. PRODUCT ID WAS PROVIDED FOR THREE SCREWS HOWEVER, IT IS UNKNOWN WHICH OF THE THREE SCREWS HAS MIGRATED. THEREFORE, THE MIGRATED SCREW COULD BE ANY OF THE FOLLOWING: 47248604840 ¿ BLUNT TIP SCREW ¿ 3181310. UDI:(B)(4). MANUFACTURING DATE: NOV 06, 2023. EXPIRATION DATE: NOV 06, 2028. 47248605040 ¿ BLUNT TIP SCREW ¿ 3174978. UDI: (B)(4). MANUFACTURING DATE: OCT 06, 2023. EXPIRATION DATE: OCT 06, 2028. 47248605240 ¿ BLUNT TIP SCREW ¿ 3174980. UDI: (B)(4). MANUFACTURING DATE: JUN 17, 2023. EXPIRATION DATE: JUN 17, 2028. 47248605440 ¿ BLUNT TIP SCREW ¿ 3163116. UDI:(B)(4). MANUFACTURING DATE: AUG 16, 2023. EXPIRATION DATE: AUG 16, 2028. D10. PROXIMAL HUMERUS, RIGHT, LONG, ÿ 10X260MM ITEM# 47249626009 LOT# 3175013. CORTICAL BONE SCREW, ÿ 4X32MM ITEM# 47248613240 LOT# 3181926. CORTICAL BONE SCREW, ÿ 4X32MM ITEM# 47248613240 LOT# 3145997. PROXIMAL HUMERUS NAIL CAP, 0MM ITEM# 47248801000 LOT# 3189241. G2. REPORT SOURCE: JAPAN. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS; HOWEVER, AN INVESTIGATION OF THE REPORTED EVENT IS IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETED, A SUPPLEMENTAL MEDWATCH 3500A WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. THE FOLLOWING SECTIONS WERE UPDATED/CORRECTED: B4, G3, G6, H2, H3, H6, H11. NO PRODUCT WAS RETURNED, OR PICTURES PROVIDED; VISUAL AND DIMENSIONAL EVALUATIONS COULD NOT BE PERFORMED. A REVIEW OF THE DEVICE MANUFACTURING RECORDS CONFIRMED NO ABNORMALITIES OR DEVIATIONS THAT COULD BE RELATED TO THE REPORTED EVENT. REVIEW OF COMPLAINT HISTORIES IDENTIFIED ADDITIONAL SIMILAR COMPLAINTS FOR THE REPORTED SCREWS AND NO ADDITIONAL SIMILAR COMPLAINTS FOR THE REPORTED PART AND LOT SCREWS COMBINATIONS. COMPLAINTS ARE MONITORED THROUGH MONTHLY COMPLAINT REVIEW IN ORDER TO IDENTIFY POTENTIAL ADVERSE TRENDS. IT CAN BE ASSUMED THAT MULTIPLE CONTRIBUTING FACTORS, RELATED TO EITHER PATIENT CONDITION AND BEHAVIOR OR IMPLANTATION PROCEDURE, CONTRIBUTED TO THE MIGRATION OF THE SCREWS. IF AND TO WHAT EXTENT ANY OF THESE ASPECTS MAY HAVE INFLUENCED THE MIGRATION OF THE SCREW REMAINS UNKNOWN. AN ADDITIONAL DEEPER INVESTIGATION WAS PERFORMED WHICH IDENTIFIED THE DESIGN LIMITATION OF THE CORELOCK MECHANISM OF THE AFFIXUS NAIL AS A POTENTIAL CONTRIBUTING FACTOR. AS PART OF THE DEEPER INVESTIGATION, A SCIENTIFIC LITERATURE SEARCH WAS CONDUCTED AND A SYSTEMATIC REVIEW OF THE OUTCOME OF INTRAMEDULLARY NAILING FOR ACUTE PROXIMAL HUMERUS FRACTURE SHOWED THAT SCREW MIGRATION AND PERFORATION INTO THE JOINT IS THE SECOND MOST COMMON COMPLICATION FOLLOWING SECONDARY LOSS OF REDUCTION. IN ADDITION, SCIENTIFIC LITERATURE OF COMPETITOR AND PREDICATE DEVICES WERE ASSESSED. THIS REVIEW HAS SHOWN THAT THE AFFIXUS® NATURAL NAIL® PROXIMAL HUMERAL SYSTEM PERFORMS BETTER AND/OR IN EQUAL RANGE IN COMPARISON WITH LEGALLY MARKETED SIMILAR DEVICES. IN CONCLUSION, AS THE CAUSE MAY BE MULTIFACTORIAL AN EXACT ROOT CAUSE COULD NOT BE IDENTIFIED FOR THE MIGRATION OF THE SCREW. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT OPERATION WAS PERFORMED WITH ANN NAIL. APPROXIMATELY ONE MONTH POST-IMPLANTATION, THE PATIENT COMPLAINED OF PAIN AND THE SURGEON FOUND #3 PROXIMAL SCREW WAS BACKED OUT FROM THE PROPER POSITION. A REVISION SURGERY WAS PERFORMED TO REMOVE THE MIGRATED SCREW. DILIGENCE IS COMPLETED AND ADDITIONAL INFORMATION IS UNAVAILABLE AT THIS TIME.

Description of Event or Problem · 0

NO ADDITIONAL EVENT INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
860823 BLUNT TIP SCREW, ÿ 4X48MM PROSTHESIS, TRAUMA HSB ZIMMER GMBH N/A 3174924 00889024505483

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Hospitalization| R