FDA Adverse Event Malfunction Summary report: N

SYRINGE 5ML SALINE FILL CHINA SP

MDR report key: 19351125 · Received May 20, 2024

Report

Report Number
2243072-2024-00635
Event Type
Malfunction
Date Received
May 20, 2024
Date of Event
April 23, 2024
Report Date
May 31, 2024
Manufacturer
BECTON DICKINSON
Product Code
NGT
UDI-DI
30382903065944
PMA / PMN Number
K161552
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

FOLLOW UP MDR FOR CORRECTION. FOLLOWING THE SUBMISSION OF THE INITIAL MDR, IT WAS NOTED THAT THE MATERIAL NUMBER WAS PROVIDED AND WAS NOT UNKNOWN AS PREVIOUSLY REPORTED. SECTION D UPDATED WITH THE CORRECT MATERIAL NUMBER AND MEDICAL DEVICE BRAND NAME SECTION G UPDATED WITH 510K.

Additional Manufacturer Narrative · 0

H.3. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED. IN THIS MDR, BD FRANKLIN LAKES, NJ HAS BEEN LISTED IN SECTIONS D.3. AND G.1. AS THE MANUFACTURING SITE IS UNKNOWN. D. THE LOT NUMBER 3090844 PROVIDED HAS NOT BEEN VERIFIED; NO REPORTED MATERIAL NUMBER.

Additional Manufacturer Narrative · 0

(B)(4), FOLLOW UP A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 306594 AND LOT NUMBER 3090844. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED. SHOULD YOU AGAIN EXPERIENCE ANY PROBLEMS WITH OUR PRODUCT WE WOULD APPRECIATE THE OPPORTUNITY TO CONDUCT A THOROUGH ANALYSIS. THERE ARE QUALITY CONTROLS CURRENTLY IN PLACE TO DETECT THIS TYPE OF DEFECT DURING THE PRODUCTION PROCESS. FURTHER ACTION HAS NOT BEEN DETERMINED NECESSARY AT THIS TIME.

Description of Event or Problem · 0

MATERIAL NUMBER PROVIDED.

Description of Event or Problem · 0

IT WAS REPORTED THAT UNDEFINED BD 5ML PREFILLED SYRINGE TIP COVER FELL OFF. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE TO ENGLISH: ON (B)(6) 2024, THE CHILD WAS ADMITTED TO THE HOSPITAL DUE TO BRONCHOPNEUMONIA, AND THE NEEDLE PLUG FELL OFF WHEN THE INFUSION TUBE WAS SEALED, AND THE NEW PREFILLED CATHETER IRRIGATOR WAS IMMEDIATELY REPLACED FOR SEALING, WHICH DID NOT CAUSE ADVERSE CONSEQUENCES.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION RECEIVED ON (B)(6) 2024, THE CHILD WAS ADMITTED TO THE HOSPITAL DUE TO BRONCHOPNEUMONIA, AND THE NEEDLE PLUG FELL OFF WHEN THE INFUSION TUBE WAS SEALED, AND THE NEW PREFILLED CATHETER IRRIGATOR WAS IMMEDIATELY REPLACED FOR SEALING, WHICH DID NOT CAUSE ADVERSE CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1147538 SYRINGE 5ML SALINE FILL CHINA SP SALINE VASCULAR ACCESS FLUSH NGT BECTON DICKINSON 3090844 30382903065944

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown