FDA Adverse Event
Malfunction
Summary report: N
PUMP MMT-722NAS PRDGM INS V2.2 SK EN
MDR report key: 1935111
·
Received December 6, 2010
Report
- Report Number
- 3004209178-2010-83512
- Event Type
- Malfunction
- Date Received
- December 6, 2010
- Date of Event
- November 23, 2010
- Report Date
- November 23, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS MED-REL
- Product Code
- MDS
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP FAILED THE DISPLACEMENT TEST AND ALARMED MOTOR ERROR DURING REWIND AS A RESULT OF THE MOTOR ENCODER SIGNAL BEING OUT OF PHASE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. TROUBLESHOOTING WAS PERFORMED. IT WAS STATED THAT THE CUSTOMER WAS ABLE TO REWIND. RAN A DISPLACEMENT TEST AND THE DEVICE FAILED THE TEST. ADVISED THE CUSTOMER TO DISCONTINUE USE OF THE DEVICE AND TO REVERT TO BACK UP PLAN. NO FURTHER INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PUMP MMT-722NAS PRDGM INS V2.2 SK EN | INFUSION PUMP | MDS | MEDTRONIC PUERTO RICO OPERATIONS MED-REL | MMT-722NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 39 YR |