FDA Adverse Event Malfunction Summary report: N

PUMP MMT-722NAS PRDGM INS V2.2 SK EN

MDR report key: 1935111 · Received December 6, 2010

Report

Report Number
3004209178-2010-83512
Event Type
Malfunction
Date Received
December 6, 2010
Date of Event
November 23, 2010
Report Date
November 23, 2010
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP FAILED THE DISPLACEMENT TEST AND ALARMED MOTOR ERROR DURING REWIND AS A RESULT OF THE MOTOR ENCODER SIGNAL BEING OUT OF PHASE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. TROUBLESHOOTING WAS PERFORMED. IT WAS STATED THAT THE CUSTOMER WAS ABLE TO REWIND. RAN A DISPLACEMENT TEST AND THE DEVICE FAILED THE TEST. ADVISED THE CUSTOMER TO DISCONTINUE USE OF THE DEVICE AND TO REVERT TO BACK UP PLAN. NO FURTHER INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-722NAS PRDGM INS V2.2 SK EN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 39 YR