FDA Adverse Event Malfunction Summary report: N

CONTOUR TEST STRIPS

MDR report key: 1935102 · Received December 6, 2010

Report

Report Number
1826988-2010-00817
Event Type
Malfunction
Date Received
December 6, 2010
Date of Event
November 24, 2010
Report Date
November 24, 2010
Manufacturer
BAYER HEALTHCARE, LLC
Product Code
NBW
PMA / PMN Number
K062058
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

THE CUSTOMER CALLED FOR HELP WITH HIS CONTOUR METER. HE PERFORMED CONTROL TESTS DURING THE CALL AND RECEIVED RESULTS OF 58, 50 AND 51 MG/DL. THE NORMAL CONTROL RANGE WAS NOT PROVIDED, BUT SHOULD BE AROUND 106-147 MG/DL. NO ADVERSE EVENTS WERE ALLEGED. THE TEST STRIPS ARE TO BE RETURNED FOR EVALUATION. REPLACEMENT TEST STRIPS AND A NEW METER WERE SENT TO THE CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTOUR TEST STRIPS NBW BAYER HEALTHCARE, LLC 7080G 9LC3C08

Patients

Seq Age Sex Outcome Treatment
1 UNK