PRECISION XTRA
Report
- Report Number
- 2954323-2010-01736
- Event Type
- Malfunction
- Date Received
- December 21, 2010
- Date of Event
- November 3, 2010
- Report Date
- December 21, 2010
- Product Code
- NBW
- Removal / Correction Number
- ADC FA1197-2010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
EVALUATION RESULTS: RETAIN SAMPLES OF PRECISION TEST STRIP LOTS MANUFACTURED WITH HOT-MELT GLUE BATCHES EXHIBITED LOWER THAN EXPECTED READINGS ON CONTINUOUS STORAGE AFTER NINE MONTHS AT 30 DEGREES CELSIUS DURING ROUTINE STABILITY PERFORMANCE TESTING. THIS STABILITY ISSUE COULD LEAD TO THE GENERATION OF INCORRECTLY DEPRESSED BLOOD GLUCOSE RESULTS AND THESE ERRONEOUS RESULTS MAY BE IN THE "C", "D" OR "E" ZONES OF PARKES ERROR GRID, AND AS SUCH, HAVE THE POTENTIAL TO BE CLINICALLY SIGNIFICANT. CUSTOMERS WILL BE NOTIFIED THROUGH A REMEDIAL ACTION (B)(4).
A CUSTOMER REPORTED THE TEST DID NOT START WHEN THE TEST STRIP WAS INSERTED INTO THEIR ADC BLOOD GLUCOSE METER AFTER BLOOD WAS APPLIED. CUSTOMER REPORTED USING AN AFFECTED TEST STRIP LOT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PRECISION XTRA | BLOOD GLUCOSE MONITORING SYSTEM | NBW | 45001A670 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |