FDA Adverse Event Malfunction Summary report: N

UV-FLASH SOLUTION TRANSFER SET(SHORT) - JAPAN

MDR report key: 1934895 · Received December 21, 2010

Report

Report Number
1423500-2010-07129
Event Type
Malfunction
Date Received
December 21, 2010
Date of Event
November 25, 2010
Report Date
November 29, 2010
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
KDJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED AT BAXTER FOR EVALUATION. HOWEVER, THE EVALUATION IS NOT COMPLETE AT THIS TIME. A 510(K) NUMBER WILL NOT BE PROVIDED IN THE EMDR AS THIS PRODUCT IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE U.S. AND DOES NOT HAVE A 510(K) NUMBER. HOWEVER, IT IS BEING REPORTED BECAUSE IT IS THE SAME OR SIMILAR TO THE PRODUCT DISTRIBUTED WITHIN THE U.S. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE DEVICE EVALUATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: ONE USED SET WAS RECEIVED WITH NO CAP ON SPIKE AND NO CAP ON PATIENT CONNECTOR IN REFERENCE TO CONNECTION ISSUE. DURING VISUAL INSPECTION, THE PATIENT CONNECTOR CONTAINED WHAT APPEARED TO BE A MARK CAUSED FROM A TORQUING THE DEVICE. NO MANUFACTURING ABNORMALITIES WERE NOTED. THE SET WAS TESTED UNDERWATER AT 8 PSI WITH NO LEAK NOTED. THE COMPLAINT COULD NEITHER BE CONFIRMED IN THE LAB NOR DUPLICATED UNDER LAB CONDITIONS WITH THE RETURNED SAMPLE. THE ROOT CAUSE OF THE COMPLAINT COULD NOT BE DETERMINED. A BATCH REVIEW COULD NOT BE PERFORMED SINCE THE LOT NUMBER WAS UNKNOWN. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BAXTER TO REPORT THAT THE SET COULD NOT BE CONNECTED BY CLEAN FLASH. THE SPIKE WAS NOT CONNECTED WITH ANYTHING WHEN THE LID OF THE CLEAN FLASH WAS OPENED. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UV-FLASH SOLUTION TRANSFER SET(SHORT) - JAPAN SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE KDJ BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1