FDA Adverse Event Malfunction Summary report: N

LIGACLIP** ENDOSCOPIC

MDR report key: 1934835 · Received December 21, 2010

Report

Report Number
3005075853-2010-07320
Event Type
Malfunction
Date Received
December 21, 2010
Date of Event
November 24, 2010
Report Date
December 1, 2010
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K864102
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED, FED, AND FORMED THE REMAINING CLIPS AS INTENDED. THE DEVICE WAS FOUND TO BE FULLY FUNCTIONAL AND CONFORMING TO OUR MANUFACTURING SPECIFICATIONS. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, THE SURGEON TRIED TO SQUEEZE THE DEVICE AND THE CLIPS WOULD NOT FORM ON THE TISSUE. THIS HAPPENED ABOUT 6 TO 8 TIMES. THE CLIPS FELL OFF THE TISSUE AND A GRASPER WAS USED TO REMOVE THE CLIPS. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO ADVERSE CONSEQUENCES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIGACLIP** ENDOSCOPIC CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1