QUICK SET PARADIGM
Report
- Report Number
- 3003442380-2024-02059
- Event Type
- Injury
- Date Received
- May 17, 2024
- Report Date
- April 25, 2025
- Product Code
- FPA
- UDI-DI
- 05705244017429
- PMA / PMN Number
- K160648
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
SUPPLEMENTAL REPORT 01 - (B)(4) - MDR 3003442380-2024-02059. CORRECTION: THIS MDR IS BEING SUBMITTED TO CORRECT THE SUBMITTED EXPIRATION DATE UNDER D4 AND MANUFACTURING DATE UNDER H4. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. THE INFORMATION IN THIS COMPLAINT (B)(4) HAS BEEN EVALUATED FOR THE MALFUNCTION CODE NO MALFUNCTION BASED ON COMPLAINT INFORMATION. THE BATCH 6003556 IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. COMPLAINT INVESTIGATION: THE REFERENCE SAMPLES CANNOT BE TESTED BECAUSE THERE WAS NO SPECIFIC MALFUNCTION TO INVESTIGATE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6003556 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 77 MANUFACTURED IN THE LINE MULTIVAC 12, ON 11/OCT/2023, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WERE IDENTIFIED RELATED TO THE MALFUNCTION REPORTED, NO MAINTENANCE EVENTS WERE RECORDED. TRENDING: A QUERY WAS RUN IN DATABASE ON 25/APR/2025 AGAINST MALFUNCTION CODE NO MALFUNCTION BASED ON COMPLAINT INFORMATION, HARM CODE UNTREATED DIABETIC KETOACIDOSIS WHICH THE PATIENT IS UNABLE TO SELF-MANAGE REQUIRING INTERVENTION BY AN HCP OR REQUIRES EMERGENCY ADVANCED LIFE SUPPORT TO PREVENT PERMANENT ORGAN DAMAGE (ELEVATED BLOOD GLUCOSE LEVEL AND SYMPTOMS E.G., MODERATE TO LARGE KETONES, NAUSEA, VOMITING, ABDOMINAL PAIN, CONFUSION) AND LOT 6003556 AND ANOTHER 1 COMPLAINT HAVE BEEN REGISTERED IN DATABASE FOR THE SAME LOT NUMBER, HARM CODE AND MALFUNCTION CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, ONLY ONE COMPLAINT RECEIVED ON THE LOT IN QUESTION, HARM CODE AND MALFUNCTION CODE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION OR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.
UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN CANADA IT WAS REPORTED THAT THE PATIENT EXPERIENCED HIGH BLOOD GLUCOSE (BG) LEVELS AND WAS HOSPITALIZED FOR A WEEK. THEIR BG LEVEL AT THE TIME OF ADMISSION WAS OVER 21 MMOL/L. THE HOSPITALIZATION WAS DUE TO HIGH BG AND DIABETIC KETOACIDOSIS (DKA). THE PATIENT WAS TAKEN TO THE HOSPITAL BY AMBULANCE AND REPORTED SYMPTOMS OF FEELING VERY SICK AND VOMITING BLOOD. THEY TESTED POSITIVE FOR KETONES, CONFIRMING THE PRESENCE OF DKA. THE PATIENT RECEIVED TREATMENT WITH AN INTRAVENOUS (IV) INSULIN DRIP DURING THEIR HOSPITAL STAY. NO FURTHER INFORMATION AVAILABLE.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 401563 | QUICK SET PARADIGM | UNO QUICK-SET 60/6 SC1 MCAN | FPA | 6003556 | 05705244017429 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Hospitalization |