UNICEL® DXC 600 PRO SYNCHRON ® CHEMISTRY ANALYZER

MDR report key: 1934645 · Received December 21, 2010

Report

Report Number
2050012-2010-01665
Event Type
Malfunction
Date Received
December 21, 2010
Date of Event
November 27, 2010
Report Date
November 27, 2010
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CUSTOMER TECHNICAL SUPPORT ASSISTED THE CUSTOMER IN CHECKING THE WASTE SUMP CANISTER. THE CUSTOMER FOUND A LOOSE NO FOAM TUBING. THE CUSTOMER TIGHTENED THE TUBING. NO FURTHER ISSUES WERE REPORTED. SERVICE WAS NOT DISPATCHED.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO THE WASTE EXIT SUMP NOT DRAINING. NO INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 600 PRO SYNCHRON ® CHEMISTRY ANALYZER CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. DXC 600 PRO N/A