FDA Adverse Event
Malfunction
Summary report: N
UNICEL® DXC 600 PRO SYNCHRON ® CHEMISTRY ANALYZER
MDR report key: 1934645
·
Received December 21, 2010
Report
- Report Number
- 2050012-2010-01665
- Event Type
- Malfunction
- Date Received
- December 21, 2010
- Date of Event
- November 27, 2010
- Report Date
- November 27, 2010
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
CUSTOMER TECHNICAL SUPPORT ASSISTED THE CUSTOMER IN CHECKING THE WASTE SUMP CANISTER. THE CUSTOMER FOUND A LOOSE NO FOAM TUBING. THE CUSTOMER TIGHTENED THE TUBING. NO FURTHER ISSUES WERE REPORTED. SERVICE WAS NOT DISPATCHED.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO THE WASTE EXIT SUMP NOT DRAINING. NO INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 600 PRO SYNCHRON ® CHEMISTRY ANALYZER | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | DXC 600 PRO | N/A |