FDA Adverse Event Injury Summary report: N

NV GXL LINER NEUTRAL, 32MM ID GROUP 1 CUPS

MDR report key: 19344192 · Received May 17, 2024

Report

Report Number
1038671-2024-01321
Event Type
Injury
Date Received
May 17, 2024
Date of Event
April 23, 2024
Report Date
September 3, 2024
Manufacturer
EXACTECH, INC.
Product Code
JDI
UDI-DI
10885862207074
PMA / PMN Number
K121392
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(H3) BASED ON THE AVAILABLE INFORMATION, THE PATIENT INVOLVED MEETS THE FOLLOWING RISK CRITERIA FOR EARLY PROSTHESIS WEAR AND/OR OSTEOLYSIS AS SPECIFIED IN THE HHE: COMBINATION OF LARGEST AVAILABLE FEMORAL HEAD AND/OR THINNEST AVAILABLE ACETABULAR LINER WAS USED AND IMPLANTED WITH A COMPONENT HAVING A SHELF AGE OF GREATER THAN 2 YEARS. THE MOST LIKELY CAUSE FOR THE REPORTED REVISION DUE TO PROSTHESIS WEAR IS A COMBINATION OF THE RISK FACTORS SPECIFIED IN AN HHE. HOWEVER, THIS CANNOT BE CONFIRMED AS THE DEVICES WERE NOT RETURNED FOR EVALUATION, AND IMAGES AND RADIOGRAPHS WERE NOT PROVIDED. *THE FOLLOWING SECTIONS HAVE CORRECTED INFORMATION: (H6) HEALTH EFFECT - CLINICAL CODE.

Additional Manufacturer Narrative · 0

PENDING INVESTIGATION. D10: 3536183 - 101-45-20 - 4.5MM DRILL BIT 20MM. 4644738 - 188-01-05 - WEDGE PLASMA X/O SZ 5. 4901074 - 170-32-00 - BIOLOX DELTA FEMORAL HEAD 32MM OD, +0MM 5140237 - 130-32-51 - NV GXL LINER. NEUTRAL, 32MM ID GROUP 1 CUPS 5976951 - 180-01-50 - NV CROWN CUP CLSTR HOLE 50MM GROUP 1. S038559 - 180-65-25 - ALTEON 6.5MM SCREW, 25MM.

Description of Event or Problem · 0

IT WAS REPORTED THAT THIS FEMALE PATIENT'S LEFT HIP WAS REVISED APPROXIMATELY 4 YEARS 4 MONTHS POST OP DUE TO EARLY POLY WEAR. REVISED TO CROWN XLE POLY. FEMALE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. UNABLE TO OBTAIN PHOTOS/X-RAYS. PRODUCT NOT RETURNING - FACILITY WOULD NOT ALLOW.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
742411 NV GXL LINER NEUTRAL, 32MM ID GROUP 1 CUPS PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED JDI EXACTECH, INC. 10885862207074

Patients

Seq Age Sex Outcome Treatment
1 NA Female Required Intervention SEE H10.