FDA Adverse Event Injury Summary report: N

BD PRESET¿ ECLIPSE¿

MDR report key: 19343554 · Received May 17, 2024

Report

Report Number
9617032-2024-00775
Event Type
Injury
Date Received
May 17, 2024
Date of Event
April 18, 2024
Report Date
May 22, 2024
Manufacturer
BECTON, DICKINSON AND COMPANY (BD)
Product Code
JKA
UDI-DI
50382903643900
PMA / PMN Number
K202446
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INITIAL REPORTER PHONE #: (B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H.6 INVESTIGATION SUMMARY MATERIAL #: 364390. LOT/BATCH #: 3081043. BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, 10 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY FUNCTIONAL TESTING AND NO ISSUES WERE OBSERVED RELATING TO DEFECTIVE LOCKING MECHANISM AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE DEFECTIVE LOCKING MECHANISM. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING THE BD PRESET¿ ECLIPSE¿ A NURSE PRICKED HER LEFT HAND DUE TO THE SAFETY SHIELD FAILING TO COVER THE IV NEEDLE. REPORT 2 OF 2.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING THE BD PRESET¿ ECLIPSE¿ A NURSE PRICKED HER LEFT HAND DUE TO THE SAFETY SHIELD FAILING TO COVER THE IV NEEDLE. REPORT 2 OF 2.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2490533 BD PRESET¿ ECLIPSE¿ BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON AND COMPANY (BD) 3081043 50382903643900

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other