BD PRESET¿ ECLIPSE¿
Report
- Report Number
- 9617032-2024-00776
- Event Type
- Injury
- Date Received
- May 17, 2024
- Date of Event
- April 18, 2024
- Report Date
- May 22, 2024
- Manufacturer
- BECTON, DICKINSON AND COMPANY (BD)
- Product Code
- JKA
- UDI-DI
- 50382903643900
- PMA / PMN Number
- K202446
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- NURSE
- Health Professional
- Yes
Narratives
INITIAL REPORTER PHONE #: (B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
H.6 INVESTIGATION SUMMARY : MATERIAL #: 364390. LOT/BATCH #: 3081043. BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, 10 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY FUNCTIONAL TESTING AND NO ISSUES WERE OBSERVED RELATING TO DEFECTIVE LOCKING MECHANISM AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE DEFECTIVE LOCKING MECHANISM. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
IT WAS REPORTED THAT WHILE USING THE BD PRESET¿ ECLIPSE¿ A NURSE PRICKED HER LEFT HAND DUE TO THE SAFETY SHIELD FAILING TO COVER THE IV NEEDLE. REPORT 1 OF 2
IT WAS REPORTED THAT WHILE USING THE BD PRESET¿ ECLIPSE¿ A NURSE PRICKED HER LEFT HAND DUE TO THE SAFETY SHIELD FAILING TO COVER THE IV NEEDLE. REPORT 1 OF 2.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2490531 | BD PRESET¿ ECLIPSE¿ | BLOOD SPECIMEN COLLECTION DEVICE | JKA | BECTON, DICKINSON AND COMPANY (BD) | 3081043 | 50382903643900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Other |