FDA Adverse Event Malfunction Summary report: N

HEARTSTRING III PROXIMAL SEAL SYSTEM, 3.8 MM

MDR report key: 1934337 · Received November 24, 2010

Report

Report Number
2242352-2010-03425
Event Type
Malfunction
Date Received
November 24, 2010
Date of Event
October 12, 2010
Report Date
October 26, 2010
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
DXC
PMA / PMN Number
K080169
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION SUMMARY: VISUAL INSPECTION REVEALED THAT THE DELIVERY TUBE WAS RETURNED SEPARATE FROM THE LOADING DEVICE. THE SEAL AND THE TENSION SPRING ASSEMBLY WERE INSIDE THE BODY OF THE LOADING DEVICE. THE GREEN SLIDE LOCK AND THE WHITE PLUNGER WERE NOT DEPRESSED ON THE DELIVERY DEVICE. THERE WAS NO EVIDENCE OF BLOOD ON THE DEVICE. BASED UPON THE RECEIVED CONDITION OF THE DEVICE, THE REPORTED COMPLAINT "SEAL WOULD NOT LOAD PROPERLY" WAS CONFIRMED. A LOT HISTORY RECORD REVIEW COULD NOT BE COMPLETED AS THE PRODUCT LOT NUMBER IS UNK. INTERNAL FILE NUMBER - (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE, THE HEARTSTRING III PROXIMAL SEAL WOULD NOT LOAD PROPERLY. A NEW KIT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PT EFFECTS. THE CORRECT LOT NUMBER COULD NOT BE OBTAINED. THE PRODUCT WAS RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HEARTSTRING III PROXIMAL SEAL SYSTEM, 3.8 MM PROXIMAL SEAL DXC MAQUET CARDIOVASCULAR, LLC HSK-3038 NI

Patients

Seq Age Sex Outcome Treatment
1 NA