FDA Adverse Event Malfunction Summary report: N

HEARTSTRING III PROXIMAL SEAL SYSTEM, 3.8 MM

MDR report key: 1934331 · Received November 24, 2010

Report

Report Number
2242352-2010-03446
Event Type
Malfunction
Date Received
November 24, 2010
Date of Event
October 25, 2010
Report Date
October 29, 2010
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
DXC
PMA / PMN Number
K080169
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION SUMMARY: VISUAL INSPECTION REVEALED THAT THE DELIVERY TUBE WAS RETURNED SEPARATE FROM THE LOADING DEVICE. THE SEAL AND THE TENSION SPRING ASSEMBLY WERE INSIDE THE BODY OF THE LOADING DEVICE. THE GREEN SLIDE LOCK AND THE WHITE PLUNGER WERE DEPRESSED ON THE DELIVERY DEVICE. THERE WAS NO EVIDENCE OF BLOOD ON THE DEVICE. BASED UPON THE RECEIVED CONDITION OF THE DEVICE, THE REPORTED COMPLAINT "SEAL BROKE" COULD NOT BE CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE PRODUCT. THERE WAS NO NONCONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE, THE HEARTSTRING III PROXIMAL SEAL BROKE. THE REPORTER STATED THAT NO OTHER INFO IS AVAILABLE REGARDING HOW THE PRODUCT BROKE. A NEW KIT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PT EFFECTS. THE PRODUCT WAS RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HEARTSTRING III PROXIMAL SEAL SYSTEM, 3.8 MM PROXIMAL SEAL DXC MAQUET CARDIOVASCULAR, LLC HSK-3038 25007183

Patients

Seq Age Sex Outcome Treatment
1 NA