ARCHITECT I2000 ANALYZER
Report
- Report Number
- 1628664-2010-00497
- Event Type
- Malfunction
- Date Received
- December 21, 2010
- Date of Event
- December 7, 2010
- Report Date
- December 7, 2010
- Manufacturer
- ABBOTT MANUFACTURING, INC.
- Product Code
- JJE
- PMA / PMN Number
- K983212
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4). AN INVESTIGATION WAS CONDUCTED TO EVALUATE THIS ISSUE. THE EVALUATION INCLUDED A REVIEW OF THE INFORMATION PROVIDED BY THE CUSTOMER, A REVIEW OF THIS INSTRUMENT ((B)(4)) SERVICE HISTORY, A REVIEW OF THE ARCHITECT SYSTEM OPERATIONS MANUAL AND A REVIEW OF THE ARCHITECT CMV IGG AVIDITY ASSAY PACKAGE INSERT. NO SPECIFIC CAUSE WAS IDENTIFIED BEHIND THIS ISSUE. HOWEVER, REVIEWING THE QUALITY CONTROLS REPORTS REVEALED THAT THE QUALITY CONTROL LEVEL NAMED CONT 1 HAD 2 RESULTS THAT WERE FLAGGED WITH THE (B)(6). THE OPERATIONS MANUAL LISTED MULTIPLE CAUSES ALONG WITH THE CORRECTIVE ACTIONS FOR ERRATIC ASSAY RESULTS. NO ADVERSE TRENDS WERE IDENTIFIED RELATED TO THIS ISSUE. THE INSTRUMENT SERVICE HISTORY REVIEW SHOWED THAT A SUBSEQUENT VISIT BY THE FIELD SERVICE ADDRESSED THE ISSUE AND ENDED UP WITH THE FIELD SERVICE REPLACING SEVERAL INSTRUMENT HARDWARE COMPONENTS INCLUDING THE SAMPLE, STAT, REAGENT 1 AND REAGENT 2 PROBES. NO PRODUCT DEFICIENCY WAS IDENTIFIED BEHIND THIS ISSUE.
(B)(4). AN INVESTIGATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.
THE CUSTOMER STATED THAT ONE PATIENT SAMPLE GENERATED A DISCREPANT RESULT FOR THE ARCHITECT CMV-IGG AVIDITY INDEX. A RESULT OF 66.2% WAS REPEATED AT 31% USING A DIFFERENT NON-ABBOTT METHOD (VIDAS). THE PATIENT WAS RETESTED FOR THE CMV AVIDITY ON THE ARCHITECT ANALYZER WITH A RESULT OF 66.6%. THE PATIENT WAS THEN TESTED FOR THE ARCHITECT CMV-IGM WITH A RESULT OF 14.71 S/CO. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ARCHITECT I2000 ANALYZER | AUTOMATED IMMUNOASSAY ANALYZER | JJE | ABBOTT MANUFACTURING, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | CMV IGG AVIDITY| CMV IGG AVIDITY |