FDA Adverse Event Injury Summary report: N

OPTI-FREE EXPRESS MULTI-PURPOSE DISINFECTING SOLUTION

MDR report key: 19341626 · Received May 17, 2024

Report

Report Number
1610287-2024-00020
Event Type
Injury
Date Received
May 17, 2024
Date of Event
April 22, 2024
Report Date
August 13, 2024
Manufacturer
ALCON RESEARCH, LLC
Product Code
LPN
PMA / PMN Number
K021143
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.3., H.6.: THE COMPLAINT SAMPLE HAS NOT RETURNED FOR EVALUATION; THE LOT NUMBER IS UNKNOWN. INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. ALICE Y. MATOBA, MD,JEFF R. PETERSON, MD, KIRK R. WILHELMUS. DENDRITIFORM KERATOPATHY ASSOCIATED WITH EXPOSURE TO POLYQUARTERNIUM-1, A COMMON OPHTHALMIC PRESERVATIVE. AMERICAN ACADEMY OF OPHTHALMOLOGY 2016; 123, 451-456. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Additional Manufacturer Narrative · 0

THE LOT NUMBER WAS NOT PROVIDED, AND THE COMPLAINT SAMPLE WAS NOT MADE AVAILABLE FOR EVALUATION. THE MANUFACTURING REVIEW DID NOT INDICATE THAT THIS COMPLAINT WAS DUE TO THE MANUFACTURING PROCESS. NO COMPLAINT OR MANUFACTURING TREND WAS IDENTIFIED. THE ROOT CAUSE COULD NOT BE DETERMINED. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Description of Event or Problem · 0

AS REPORTED IN THE LITERATURE OUT OF SIXTEEN CONSUMERS, TWELVE CONSUMERS WERE EXPOSURE TO CONTACT LENS CARE DISINFECTING SOLUTION CONTAINING POLYQUATERNIUM-1. THE ACTUAL CONTACT LENS CARE DISINFECTING SOLUTIONS ASSOCIATED WITH THE EVENT IS NOT KNOWN, THEREFORE AS PER THE ARTICLE UNSPECIFIED CONTACT LENS CARE DISINFECTING SOLUTIONS CONSIDERED AS A SUSPECT PRODUCT. THIS COMPLAINT REFERS TO ONE OF THE CONSUMERS OUT OF TWELVE, WHO WAS EXPOSED CONTACT LENS CARE DISINFECTING SOLUTION CONTAINING POLYQUATERNIUM-1 AND WAS IN USE WITH DAILY CONTACT LENS FOR BOTH EYES, INITIALLY IT WAS MISDIAGNOSED AS HERPEESE SIMPLEX VIRUS AND ACANTHAMOEBA KERATITIS. HOWEVER, WAS CORRECTLY DIAGNOSED AS DENDRITIFORM KERATOPATHY ASSOCIATED WITH EXPOSURE TO POLYQUATERNIUM-1. CONTACT LENS SOLUTION WAS DISCONTINUED. THE STATUS OF THE CONSUMER'S EYE WAS RESOLVED (FOUR WEEKS OF DURATION) AT THE TIME OF THIS REPORT. NO ADDITIONAL INFORMATION REQUESTED DUE TO THE UNLIKELIHOOD OF OBTAINING ADDITIONAL INFORMATION FROM LITERATURE ARTICLES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
905946 OPTI-FREE EXPRESS MULTI-PURPOSE DISINFECTING SOLUTION ACCESSORIES, SOFT LENS PRODUCTS LPN ALCON RESEARCH, LLC NA ASKU

Patients

Seq Age Sex Outcome Treatment
1 57 YR Female Other UNSPECIFIED CONTACT LENS