FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 1934115 · Received December 2, 2010

Report

Report Number
2531779-2010-02680
Event Type
Malfunction
Date Received
December 2, 2010
Report Date
November 3, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARED TO BE INTACT WITH NO LIFTING OR PEELING OBSERVED, THE UP, DOWN AND OK KEYPAD BUTTONS INTERMITTENTLY RESPONDED TO AND REQUIRED EXCESSIVE FORCE TO ACTIVATE DURING TESTING, THE CONTRAST KEYPAD BUTTON WAS UNRESPONSIVE TO USER INPUT, IT WAS OBSERVED THAT THE UP AND OK KEY CONTACTS WERE MISALIGNED, THE CONTRAST KEY CONTACT WAS INVERTED AND DID NOT ALWAYS SPRING BACK, THE UP, DOWN AND OK KEY CONTACT BECAME INVERTED WHEN PRESSED AND DID NOT ALWAYS SPRING BACK AND THERE WAS EVIDENCE OF ADHESIVE/CONTAMINATION UNDER ALL THE KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE UP ARROW KEYPAD BUTTON SLOW TO RESPOND WHEN PRESSED. CLAIMED KEYPAD IS INTACT. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment
1