FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 1934085 · Received December 2, 2010

Report

Report Number
2531779-2010-02683
Event Type
Malfunction
Date Received
December 2, 2010
Report Date
November 5, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARS TO BE SWOLLEN. THE PUMP WAS RETURNED WITH THE KEYPAD LIFTING UP BELOW THE "OK" BUTTON. THE "UP/DOWN/OK" KEYPAD BUTTON IS INTERMITTENTLY RESPONDING DURING TESTING. THE "CONTRAST" KEYPAD BUTTON REQUIRE EXCESSIVE FORCE TO ACTIVATE. THE KEYPAD WAS REMOVED FOR FURTHER INVESTIGATION. DURING A VISION INSPECTION, THE "CONTRAST" KEY CONTACT IS INVERTED AND DO NOT ALWAYS SPRING BACK, AND THE "UP/DOWN/OK" KEY CONTACTS BECOME INVERTED WHEN PRESSED AND DO NOT ALWAYS SPRING BACK. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

THE DISTRIBUTOR/REPORTER ALLEGED THAT THE UP AND DOWN BUTTONS ON THE KEYPAD DOES NOT RESPOND. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.