ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2010-02683
- Event Type
- Malfunction
- Date Received
- December 2, 2010
- Report Date
- November 5, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- UNKNOWN
Narratives
THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARS TO BE SWOLLEN. THE PUMP WAS RETURNED WITH THE KEYPAD LIFTING UP BELOW THE "OK" BUTTON. THE "UP/DOWN/OK" KEYPAD BUTTON IS INTERMITTENTLY RESPONDING DURING TESTING. THE "CONTRAST" KEYPAD BUTTON REQUIRE EXCESSIVE FORCE TO ACTIVATE. THE KEYPAD WAS REMOVED FOR FURTHER INVESTIGATION. DURING A VISION INSPECTION, THE "CONTRAST" KEY CONTACT IS INVERTED AND DO NOT ALWAYS SPRING BACK, AND THE "UP/DOWN/OK" KEY CONTACTS BECOME INVERTED WHEN PRESSED AND DO NOT ALWAYS SPRING BACK. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.
THE DISTRIBUTOR/REPORTER ALLEGED THAT THE UP AND DOWN BUTTONS ON THE KEYPAD DOES NOT RESPOND. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ANIMAS 2020 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |