FDA Adverse Event Malfunction Summary report: N

PUMP,COLLEAGUE CXE SINGLECHANNEL COLOUR-LOANER

MDR report key: 1933893 · Received December 21, 2010

Report

Report Number
6000001-2010-06065
Event Type
Malfunction
Date Received
December 21, 2010
Date of Event
June 21, 2010
Report Date
July 28, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION WAS CONFIRMED AND DUPLICATED. THE ASSIGNABLE CAUSE WERE DEPLETED MAIN BATTERIES. THE MAIN BATTERIES AND HARNESS HAVE BEEN REPLACED. ADDITIONAL: THE USER INTERFACE MODULE MASTER SOFTWARE VERSION FOR THIS DEVICE IS 6.13.92, CATEGORIZED AS REMEDIATED. A SERVICE HISTORY REVIEW REVEALED THAT THIS DEVICE WAS PREVIOUSLY SERVICED FOR THE REPORTED CONDITION. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED WHEN THE EVALUATION RESULTS OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE CONTACTED BAXTER (B)(4) TO REPORT A COLLEAGUE INFUSION PUMP THAT HAD A DAMAGED BATTERY. THERE WAS NO ADVERSE EVENT, PATIENT INJURY, OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT. DURING BAXTER'S REVIEW OF THE EVENT HISTORY, IT WAS DISCOVERED THAT THERE WAS A BATTERY DEPLETED SET ALARM THAT HAD OCCURRED DURING INFUSION WHICH CAUSED AN INTERRUPTION DURING DELIVERY. THERE IS NO FURTHER COMPLAINT INFORMATION AVAILABLE. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION IS CURRENTLY UNKNOWN FOR THIS DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP,COLLEAGUE CXE SINGLECHANNEL COLOUR-LOANER PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1