FDA Adverse Event Malfunction Summary report: N

FISHER & PAYKEL HEALTHCARE

MDR report key: 19338250 · Received May 17, 2024

Report

Report Number
9611451-2024-00326
Event Type
Malfunction
Date Received
May 17, 2024
Date of Event
April 15, 2024
Report Date
April 18, 2024
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BTT
PMA / PMN Number
K020332
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

(B)(4). CORRECTIONS: SECTION D2A: COMMON DEVICE NAME UPDATED TO TUBING KIT. SECTION G3: DATE CORRECTED. SECTION D4: THE LOT AND UDI #S OF THE RT330 OPTIFLOW TUBING KITS WERE NOT RECEIVED. SECTION H11: METHOD: THE TWO SUBJECT MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBERS THAT WERE PROVIDED AS PART OF RT330 OPTIFLOW TUBING KITS WERE RETURNED TO FISHER & PAYKEL HEALTHCARE IN NEW ZEALAND FOR EVALUATION. RESULTS: THE TWO MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBERS WERE VISUALLY INSPECTED. A HORIZONTAL CRACK AND A STRESSMARK WAS IDENTIFIED ON THE DOME OF ONE OF THE CHAMBERS. BOTH CHAMBERS WERE FOUND WITH THE WATER FEEDSET SPIKE MISSING. CONCLUSION: OUR INVESTIGATION IDENTIFIED A CRACK ON THE DOME OF ONE OF THE CHAMBERS. BOTH CHAMBERS WERE ALSO FOUND WITH THE WATER FEEDSET SPIKE MISSING. WE ARE UNABLE TO DETERMINE THE CAUSE OF THE DAMAGE. EVERY MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER IS PRESSURE TESTED AND VISUALLY INSPECTED FOLLOWING THE MANUFACTURING PROCESS AND ANY HOLES, CRACKS, OR LEAKS WOULD BE IDENTIFIED DURING THIS PROCESS. IN ADDITION, DESTRUCTIVE PULL TESTING IS PERFORMED HOURLY TO VERIFY THE STRENGTH OF THE WATER FEEDSET TUBING. ANY CHAMBER WHICH FAILS THESE TESTS IS REJECTED. THE SUBJECT MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBERS WOULD HAVE MET THE REQUIRED SPECIFICATIONS AT THE TIME OF PRODUCTION. THE USER INSTRUCTIONS THAT ACCOMPANY THE RT330 OPTIFLOW TUBING KIT STATE THE FOLLOWING: DO NOT USE THE CHAMBER IF THE SEALS ARE NOT INTACT WHEN RECEIVED, OR IF IT HAS BEEN DROPPED. ENSURE THERE IS A WATER SUPPLY CONNECTED TO THE CHAMBER AND THAT WATER IS PRESENT WITHIN THE CHAMBER. APPROPRIATE PATIENT MONITORING (E.G. OXYGEN SATURATION) MUST BE USED AT ALL TIMES. ENSURE APPROPRIATE VENTILATOR OR FLOW SOURCE ALARMS ARE SET BEFORE CONNECTING BREATHING SETS TO PATIENT. VISUALLY INSPECT BREATHING SETS FOR DAMAGE (E.G. CRUSHED TUBE OR CRACKED CONNECTOR) BEFORE USE AND REPLACE IF DAMAGED.

Additional Manufacturer Narrative · 0

(B)(4). THE TWO SUBJECT MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBERS HAVE BEEN REQUESTED TO BE RETURNED TO FISHER & PAYKEL HEALTHCARE IN NEW ZEALAND FOR EVALUATION. WE WILL PROVIDE A FOLLOW UP REPORT UPON COMPLETION OF OUR INVESTIGATION.

Additional Manufacturer Narrative · 0

(B)(4). CORRECTIONS: SECTION B4: CORRECTED YEAR. SECTION D1: CORRECTED BRAND NAME. SECTION D2A: CORRECTED COMMON DEVICE NAME SECTION D4: MODEL AND CATALOG # CORRECTED TO RT330. SECTION G4: UPDATED 510(K). SECTION D4: THE LOT AND UDI #S OF THE RT330 OPTIFLOW TUBING KITS WERE NOT RECEIVED. SECTION H11: METHOD: THE TWO SUBJECT MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBERS THAT WERE PROVIDED AS PART OF RT330 OPTIFLOW TUBING KITS WERE RETURNED TO FISHER & PAYKEL HEALTHCARE IN NEW ZEALAND FOR EVALUATION. RESULTS: THE TWO MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBERS WERE VISUALLY INSPECTED. A HORIZONTAL CRACK AND A STRESSMARK WAS IDENTIFIED ON THE DOME OF ONE OF THE CHAMBERS. BOTH CHAMBERS WERE FOUND WITH THE WATER FEEDSET SPIKE MISSING. CONCLUSION: OUR INVESTIGATION IDENTIFIED A CRACK ON THE DOME OF ONE OF THE CHAMBERS. BOTH CHAMBERS WERE ALSO FOUND WITH THE WATER FEEDSET SPIKE MISSING. WE ARE UNABLE TO DETERMINE THE CAUSE OF THE DAMAGE. EVERY MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER IS PRESSURE TESTED AND VISUALLY INSPECTED FOLLOWING THE MANUFACTURING PROCESS AND ANY HOLES, CRACKS, OR LEAKS WOULD BE IDENTIFIED DURING THIS PROCESS. IN ADDITION, DESTRUCTIVE PULL TESTING IS PERFORMED HOURLY TO VERIFY THE STRENGTH OF THE WATER FEEDSET TUBING. ANY CHAMBER WHICH FAILS THESE TESTS IS REJECTED. THE SUBJECT MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBERS WOULD HAVE MET THE REQUIRED SPECIFICATIONS AT THE TIME OF PRODUCTION. THE USER INSTRUCTIONS THAT ACCOMPANY THE RT330 OPTIFLOW TUBING KIT STATE THE FOLLOWING: DO NOT USE THE CHAMBER IF THE SEALS ARE NOT INTACT WHEN RECEIVED, OR IF IT HAS BEEN DROPPED. ENSURE THERE IS A WATER SUPPLY CONNECTED TO THE CHAMBER AND THAT WATER IS PRESENT WITHIN THE CHAMBER. APPROPRIATE PATIENT MONITORING (E.G. OXYGEN SATURATION) MUST BE USED AT ALL TIMES. ENSURE APPROPRIATE VENTILATOR OR FLOW SOURCE ALARMS ARE SET BEFORE CONNECTING BREATHING SETS TO PATIENT. VISUALLY INSPECT BREATHING SETS FOR DAMAGE (E.G. CRUSHED TUBE OR CRACKED CONNECTOR) BEFORE USE AND REPLACE IF DAMAGED.

Description of Event or Problem · 0

A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN COLUMBUS, USA, VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE, THAT TWO MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBERS, PROVIDED WITH RT330 OPTIFLOW¿ JUNIOR BREATHING CIRCUIT KITS, WERE FOUND TO BE LEAKING WATER. THERE WAS NO PATIENT HARM.

Description of Event or Problem · 0

A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN COLUMBUS, USA, VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE, THAT TWO MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBERS WERE FOUND TO BE LEAKING WATER. THERE WAS NO PATIENT HARM.

Description of Event or Problem · 0

A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN COLUMBUS, USA, VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE, THAT TWO MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBERS, PROVIDED WITH RT330 OPTIFLOW¿ JUNIOR BREATHING CIRCUIT KITS, WERE FOUND TO BE LEAKING WATER. THERE WAS NO PATIENT HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2500221 FISHER & PAYKEL HEALTHCARE TUBING KIT BTT FISHER & PAYKEL HEALTHCARE LTD RT330 SEE H11

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown F&P RT330 OPTIFLOW¿ JUNIOR BREATHING CIRCUIT KIT