FDA Adverse Event Injury Summary report: N

NEXUS® SONASTAR® SHORT HANDPIECE

MDR report key: 19337652 · Received May 16, 2024

Report

Report Number
2435119-2024-00004
Event Type
Injury
Date Received
May 16, 2024
Date of Event
April 15, 2024
Report Date
February 12, 2025
Manufacturer
MISONIX, INC.
Product Code
LFL
PMA / PMN Number
K231117
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FI
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

ON APRIL 16, 2024, MISONIX® LLC., A BIOVENTUS® CO., RECEIVED A REPORT OF AN EVENT INVOLVING A NEXUS® SONASTAR® SHORT HANDPIECE (PART NUMBER 100-24-0000, SERIAL NUMBER (B)(6), A NEXUS® SONASTAR® ASPIRATION TIP 1.6MM MICRO SHORT, + IRRIGATION & ASPIRATION (PART NUMBER 130-35-1416), AND A NEXUS® CONSOLE (PART NUMBER 100-10-0000, SERIAL NUMBER (B)(6) DURING AN OPEN LIVER RESECTION THAT OCCURRED ON (B)(6) 2024. SPECIFICALLY, IT WAS REPORTED DURING THE OPERATION THAT THE SILICONE COVER OF THE DEVICE'S HANDPIECE LEFT A BURN MARK ON THE GLOVE AND FINGER OF THE SCRUB NURSE AND OPERATING SURGEON. THE DEGREE OF THE BURN WAS NOT SPECIFIED. PERMANENT IMPAIRMENT TO BODY STRUCTURE OR BODY FUNCTION WAS NOT REPORTED. MEDICAL INTERVENTION REQUIRED TO PRECLUDE SERIOUS INJURY WAS NOT REPORTED. THE EFFECTS OF THE EVENT WERE LIMITED TO THE SCRUB NURSE AND OPERATING SURGEON. ON MAY 14, 2024, MISONIX RECEIVED ADDITIONAL DETAILS OF THE EVENT. A DELAY IN TREATMENT WAS REPORTED THAT DID NOT EXCEED 15 MINUTES WHILE THE PATIENT WAS UNDER ANESTHESIA. THE OPERATION WAS ABLE TO CONTINUE WITH THE USE OF ANOTHER DEVICE. THE THERMAL INJURY TO THE SCRUB NURSE AND OPERATING SURGEON WAS INVESTIGATED AFTER THE COMPLETION OF THE OPERATION. THE LOT NUMBER FOR THE NEXUS® SONASTAR® ASPIRATION TIP 1.6MM MICRO SHORT, + IRRIGATION & ASPIRATION (PART NUMBER 130-35-1416) WAS NOT REPORTED; THEREFORE, A COMPLETE DEVICE HISTORY RECORD REVIEW CANNOT BE PERFORMED. THE DEVICE HISTORY RECORD WAS REVIEWED FOR THE NEXUS® SONASTAR® SHORT HANDPIECE (PART NUMBER 100-24-0000, SERIAL NUMBER (B)(6) AND NEXUS® CONSOLE (PART NUMBER 100-10-0000, SERIAL NUMBER (B)(6) IN USE AT THE TIME OF THE EVENT. THERE WERE NO DEVIATIONS FOUND DURING IN-PROCESS OR FINAL INSPECTION OF THE DEVICES. INSPECTION AND TEST RESULTS MET MISONIX SPECIFICATIONS PRIOR TO RELEASE TO COMMERCE. A REVIEW OF POST MARKET SURVEILLANCE DATA FOR THE NEXUS® SYSTEM DID NOT SHOW ANY SIGNIFICANT ADVERSE TRENDS FOR THERMAL INJURY TO THE PATIENT OR USER. THE CURRENT FREQUENCY OF OCCURRENCE IS WITHIN THE FREQUENCY IN THE ORIGINAL RISK MANAGEMENT REPORT. THERE IS NO CHANGE TO THE RESIDUAL RISK OR RISK-BENEFIT RATIO OF THE DEVICE. DUE TO THE NATURE IN WHICH ULTRASONIC DEVICES TRANSFER ENERGY INTO THE TARGET TISSUE, THERMAL EVENTS MAY OCCUR. THE INSTRUCTIONS FOR USE MANUAL (100-10-1000, REVISION J) FOR THE NEXUS® ULTRASONIC SURGICAL ASPIRATION SYSTEM CONTAINS THE FOLLOWING WARNINGS AND CAUTIONS REGARDING THE SURGICAL TECHNIQUE AND DEVICE SETTINGS REQUIRED TO PREVENT THERMAL INJURY TO THE PATIENT OR USER: WARNING THE NEXUS® ULTRASONIC SURGICAL ASPIRATOR SYSTEM IS INTENDED TO BE USED IN VARIOUS TYPES OF INVASIVE, SURGICAL PROCEDURES. THERE MAY BE INDIRECT DANGER TO THE PATIENT SHOULD THE DEVICE FAIL DURING THE PROCEDURE. IT IS RECOMMENDED THAT THE FACILITY FOLLOWS ITS BACK-UP EQUIPMENT PROTOCOLS. CAUTION THE SYSTEM CHECK SHOULD ALWAYS BE DONE IN ADVANCE OF PREPARING PATIENT FOR SURGERY TO MINIMIZE RISK TO PATIENT IN CASE OF SYSTEM MALFUNCTION. CAUTION IT IS STRONGLY ADVISED THAT A STERILE BACKUP HANDPIECE BE READILY AVAILABLE IN THE OPERATING ROOM AS INSURANCE ANY CONTAMINATION OR MALFUNCTION OF THE HANDPIECE USED DURING SURGERY. POTENTIAL BURN HAZARD WARNING NEXUS® PROBES HAVE A SILICONE OR HARD PLASTIC SHEATH. COMPRESSING OR BENDING THE SHEATH MAY CAUSE THE SHEATH TO CONTACT THE VIBRATING SURFACE ALONG THE LENGTH OF THE PROBE OR AT THE PROBE TIP AND MAY CAUSE EXCESSIVE HEATING, WHICH MAY BURN USER OR PATIENT TISSUE AT THE SURGICAL SITE. WARNING EXCESSIVE LOADING OF NEXUS® PROBES AT THE SURGICAL SITE MAY INDUCE HEATING DUE TO VIBRATION AND FRICTION AS TARGET TISSUE IS FRAGMENTED AND EMULSIFIED. IT IS CRITICAL TO MANAGE THE TEMPERATURE OF THE PROBE BY ADJUSTING THE IRRIGATION, ASPIRATION, AND ULTRASOUND SETTINGS, AND SURGICAL TECHNIQUE. TISSUE NECROSIS MAY RESULT IF PROBE TIP IS NOT MOVED RELATIVE TO TISSUE. A CONTINUOUS, LATERAL SWEEPING MOTION IS RECOMMENDED IN ORDER TO MINIMIZE CONTACT DURATION WITH THE ULTRASONIC PROBE TIP AND MINIMIZE HEAT BUILD-UP. WHEN LATERAL MOTION IS NOT POSSIBLE WITHDRAW AND RE-INSERT PROBE TIP FREQUENTLY. WARNING CONTACT TO VIBRATING ELEMENTS LIKE AN EXTENSION AND ULTRASONIC PROBE TIP MAY CAUSE BURNS AND SHOULD BE AVOIDED BY ALL MEANS. THE HANDPIECE SHOULD ONLY BE HELD AT THE BLACK HANDPIECE HOUSING AREA AND/OR THE BLACK HARD SHEATH. WARNING A PROTECTIVE SILICONE SLEEVE, INCLUDED WITH CERTAIN PROBE TIPS, REDUCES THE RISK OF THERMAL DAMAGE BUT DOES NOT ELIMINATE IT. CONTACT WITH THE SILICONE SLEEVE SHOULD BE AVOIDED OR KEPT BRIEF WITH MINIMAL AMOUNT OF CONTACT PRESSURE. PRESSURE AND EXTENDED EXPOSURE CAN STILL RESULT IN EXCESSIVE FRICTIONAL HEAT AND CAUSE BURNS. WARNING CONTACT OF THE RIGID OR SILICONE SHEATHS WITH PATIENT TISSUE UNDER PRESSURE, MAY CREATE A BURN HAZARD. AVOID CONTACT OF SHEATH ELEMENTS WITH PATIENT TISSUE UNDER PRESSURE. WARNING PROBE TIP TEMPERATURES MAY EXCEED THE TISSUE NECROSIS POINT IF INSUFFICIENT IRRIGANT IS PRESENT AT THE PROBE TIP-TISSUE INTERFACE. FOR HARD TISSUE REMOVAL, ALWAYS USE THE MAXIMUM IRRIGATION FLOWRATE THAT DOES NOT AFFECT THE SURGICAL FIELD OF VIEW, OR IMPACT SURGICAL TECHNIQUE. ADDITIONAL EXTERNAL IRRIGATION, E.G., BY ADMINISTERING STERILE SALINE WITH A SYRINGE OVER THE DISTAL PROBE TIP PORTION, MAY BE NECESSARY FOR REMOVAL OF VERY DENSE, HARD OSSEOUS STRUCTURES. WARNING HARD TISSUE APPLICATIONS, A MINIMUM IRRIGATION SETTING OF 20 IS RECOMMENDED TO MINIMIZE OR PREVENT THERMAL INJURY AND/OR TISSUE NECROSIS. CAUTION INSUFFICIENT IRRIGATION AND HIGH TIP PRESSURE (LOADING) UNDER EXTENDED EXPOSURE, E.G., IN TIGHT CAVITIES, ARE TO BE AVOIDED WHILE REMOVING HARD TISSUE. IT IS RECOMMENDED TO WITHDRAW AND RE-INSERT THE ULTRASONIC TIPS (E.G., BLADES & SHAVERS) REPEATEDLY TO RE-ESTABLISH ADEQUATE COOLING AND LUBRICATION. CAUTION ADDITIONAL EXTERNAL IRRIGATION, E.G., BY ADMINISTERING STERILE SALINE WITH A SYRINGE OVER THE DISTAL TIP PORTION, MAY BE NECESSARY WHEN REMOVING VERY DENSE, HARD OSSEOUS STRUCTURES. CAUTION PRIME THE IRRIGATION TUBING PRIOR TO USE. AT ALL TIMES ENSURE THAT THE IRRIGATION FLOWS TOWARDS THE HANDPIECE WHEN FOOTSWITCH IS DEPRESSED. IF NO IRRIGATION IS FLOWING, CEASE USE UNTIL FLOW IS RESTORED. THE SUBJECT HANDPIECE, DISPOSABLE OR CONSOLE HAS NOT YET BEEN RETURNED FOR EVALUATION; THEREFORE, A THOROUGH INVESTIGATION COULD NOT BE CONDUCTED AS OF THE TIME OF THIS REPORT. MISONIX IS ACTIVELY WORKING WITH THE CUSTOMER FOR THE RETURN OF THE DEVICES USED DURING THE EVENT TO COMPLETE THE INVESTIGATION. UPON RETURN, A FULL EVALUATION WILL BE COMPLETED TO ESTABLISH POTENTIAL ROOT CAUSE.

Additional Manufacturer Narrative · 0

FOLLOW UP # 1. THE SUBJECT GENERATOR AND HANDPIECE WERE RETURNED TO MISONIX FOR EVALUATION. INSPECTION AND TESTING CONCLUDED BOTH DEVICES CONFORMED TO SPECIFICATIONS. NO MALFUNCTIONS THAT WOULD CAUSE OR CONTRIBUTE TO A USER BURN WERE NOTED. THE INVESTIGATION HAS BEEN CONCLUDED.

Description of Event or Problem · 0

ON APRIL 16, 2024, MISONIX® LLC., A BIOVENTUS® CO., RECEIVED A REPORT OF AN EVENT INVOLVING A NEXUS® SONASTAR® SHORT HANDPIECE (PART NUMBER 100-24-0000, SERIAL NUMBER (B)(6), A NEXUS® SONASTAR® ASPIRATION TIP 1.6MM MICRO SHORT, + IRRIGATION & ASPIRATION (PART NUMBER 130-35-1416), AND A NEXUS® CONSOLE (PART NUMBER 100-10-0000, SERIAL NUMBER (B)(6) DURING AN OPEN LIVER RESECTION THAT OCCURRED ON (B)(6) 2024. SPECIFICALLY, IT WAS REPORTED DURING THE OPERATION THAT THE SILICONE COVER OF THE DEVICE'S HANDPIECE LEFT A BURN MARK ON THE GLOVE AND FINGER OF THE SCRUB NURSE AND OPERATING SURGEON. THE DEGREE OF THE BURN WAS NOT SPECIFIED. PERMANENT IMPAIRMENT TO BODY STRUCTURE OR BODY FUNCTION WAS NOT REPORTED. MEDICAL INTERVENTION REQUIRED TO PRECLUDE SERIOUS INJURY WAS NOT REPORTED. THE EFFECTS OF THE EVENT WERE LIMITED TO THE SCRUB NURSE AND OPERATING SURGEON. ON MAY 14, 2024, MISONIX RECEIVED ADDITIONAL DETAILS OF THE EVENT. A DELAY IN TREATMENT WAS REPORTED THAT DID NOT EXCEED 15 MINUTES WHILE THE PATIENT WAS UNDER ANESTHESIA. THE OPERATION WAS ABLE TO CONTINUE WITH THE USE OF ANOTHER DEVICE. THE THERMAL INJURY TO THE SCRUB NURSE AND OPERATING SURGEON WAS INVESTIGATED AFTER THE COMPLETION OF THE OPERATION.

Description of Event or Problem · 0

ON APRIL 16, 2024, MISONIX® LLC., A BIOVENTUS® CO., RECEIVED A REPORT OF AN EVENT INVOLVING A NEXUS® SONASTAR® SHORT HANDPIECE (PART NUMBER 100-24-0000, SERIAL NUMBER (B)(6)), A NEXUS® SONASTAR® ASPIRATION TIP 1.6MM MICRO SHORT, + IRRIGATION & ASPIRATION (PART NUMBER 130-35-1416), AND A NEXUS® CONSOLE (PART NUMBER 100-10-0000, SERIAL NUMBER (B)(6)) DURING AN OPEN LIVER RESECTION THAT OCCURRED ON (B)(6) 2024. SPECIFICALLY, IT WAS REPORTED DURING THE OPERATION THAT THE SILICONE COVER OF THE DEVICE'S HANDPIECE LEFT A BURN MARK ON THE GLOVE AND FINGER OF THE SCRUB NURSE AND OPERATING SURGEON. THE DEGREE OF THE BURN WAS NOT SPECIFIED. PERMANENT IMPAIRMENT TO BODY STRUCTURE OR BODY FUNCTION WAS NOT REPORTED. MEDICAL INTERVENTION REQUIRED TO PRECLUDE SERIOUS INJURY WAS NOT REPORTED. THE EFFECTS OF THE EVENT WERE LIMITED TO THE SCRUB NURSE AND OPERATING SURGEON. ON MAY 14, 2024, MISONIX RECEIVED ADDITIONAL DETAILS OF THE EVENT. A DELAY IN TREATMENT WAS REPORTED THAT DID NOT EXCEED 15 MINUTES WHILE THE PATIENT WAS UNDER ANESTHESIA. THE OPERATION WAS ABLE TO CONTINUE WITH THE USE OF ANOTHER DEVICE. THE THERMAL INJURY TO THE SCRUB NURSE AND OPERATING SURGEON WAS INVESTIGATED AFTER THE COMPLETION OF THE OPERATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
971721 NEXUS® SONASTAR® SHORT HANDPIECE ULTRASONIC SURGICAL ASPIRATOR SYSTEM HANDPIECE LFL MISONIX, INC. 100-24-0000

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other