BED CHECK
Report
- Report Number
- 1929691-2010-00003
- Event Type
- Death
- Date Received
- December 20, 2010
- Date of Event
- December 6, 2010
- Report Date
- December 6, 2010
- Manufacturer
- STANLEY SECURITY SOLUTIONS, INC., SENIOR TECHNOLOGIES DIVISION
- Product Code
- KMI
- Report Source
- Manufacturer report
- Reporter Location
- NH, US
- Reporter Occupation
- OTHER
Narratives
ON 12-9-2010: THE EQUIPMENT WAS RETURNED BY (B)(6) AND WAS FULLY EVALUATED. THE BED CHECK CLASSIC-CHECK MODEL 72021 AND BED SENSORMAT 74011 TESTED ACCORDING TO MANUFACTURER SPECIFICATION. THE RESIDENT WAS CAUGHT BETWEEN THE BED RAIL AND BED MATTRESS WHICH DID NOT ALLOW FOR A FULL EXIT. AS REPORTED BY THE FACILITY THE RESIDENT REMAINED ON THE MAT, THEREFORE THE MAT DID NOT ALARM. UPON REMOVING THE RESIDENT FROM THE MAT, THE FACILITY REPORTED THAT THE BED CHECK MONITOR AND PAD IMMEDIATELY SIGNALED AN ALARM VERIFYING THE PRODUCT WAS FULLY OPERATIONAL. ADDITIONAL MODEL NUMBER: 74011, LOT# A0210 (74011).
RESIDENT WAS FOUND WITH LOWER BODY ON FLOOR AND UPPER BODY LEANING AGAINST BED. RESIDENT'S HEAD WAS ON BED ON LEFT SIDE IN BETWEEN MATTRESS AND SIDE RAIL. BED ALARM DID NOT SOUND AS HEAD WAS ON THE BED ALARM SENSOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BED CHECK | CLASSIC-CHECK AND BED SENSORMAT | KMI | STANLEY SECURITY SOLUTIONS, INC., SENIOR TECHNOLOGIES DIVISION | 72021 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 88 YR | Death |