FDA Adverse Event Death Summary report: N

BED CHECK

MDR report key: 1933744 · Received December 20, 2010

Report

Report Number
1929691-2010-00003
Event Type
Death
Date Received
December 20, 2010
Date of Event
December 6, 2010
Report Date
December 6, 2010
Manufacturer
STANLEY SECURITY SOLUTIONS, INC., SENIOR TECHNOLOGIES DIVISION
Product Code
KMI
Report Source
Manufacturer report
Reporter Location
NH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ON 12-9-2010: THE EQUIPMENT WAS RETURNED BY (B)(6) AND WAS FULLY EVALUATED. THE BED CHECK CLASSIC-CHECK MODEL 72021 AND BED SENSORMAT 74011 TESTED ACCORDING TO MANUFACTURER SPECIFICATION. THE RESIDENT WAS CAUGHT BETWEEN THE BED RAIL AND BED MATTRESS WHICH DID NOT ALLOW FOR A FULL EXIT. AS REPORTED BY THE FACILITY THE RESIDENT REMAINED ON THE MAT, THEREFORE THE MAT DID NOT ALARM. UPON REMOVING THE RESIDENT FROM THE MAT, THE FACILITY REPORTED THAT THE BED CHECK MONITOR AND PAD IMMEDIATELY SIGNALED AN ALARM VERIFYING THE PRODUCT WAS FULLY OPERATIONAL. ADDITIONAL MODEL NUMBER: 74011, LOT# A0210 (74011).

Description of Event or Problem · 1

RESIDENT WAS FOUND WITH LOWER BODY ON FLOOR AND UPPER BODY LEANING AGAINST BED. RESIDENT'S HEAD WAS ON BED ON LEFT SIDE IN BETWEEN MATTRESS AND SIDE RAIL. BED ALARM DID NOT SOUND AS HEAD WAS ON THE BED ALARM SENSOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BED CHECK CLASSIC-CHECK AND BED SENSORMAT KMI STANLEY SECURITY SOLUTIONS, INC., SENIOR TECHNOLOGIES DIVISION 72021

Patients

Seq Age Sex Outcome Treatment
1 88 YR Death