TALENT ABDOMINAL STENT GRAFT SYSTEM
Report
- Report Number
- 2953200-2010-02563
- Event Type
- Injury
- Date Received
- December 16, 2010
- Report Date
- November 17, 2010
- Manufacturer
- MEDTRONIC CARDIOVASCULAR SANTA ROSA
- Product Code
- MIH
- PMA / PMN Number
- P070027
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). EVALUATION: RESULTS: (STENT GRAFT MIGRATION, ENDOLEAK), (DISEASE PROGRESSION, ANGULATED AORTIC BIFURCATION, DILATED AND SHORT AORTIC NECK). CONCLUSION: (DISEASE PROGRESSION, ANGULATED AORTIC BIFURCATION, DILATED AND SHORT AORTIC NECK).
A TALENT STENT GRAFT WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM APPROX 7 YEARS AGO. ANEURYSM AND VESSEL MORPHOLOGY AT THE TIME OF IMPLANT WERE NOT REPORTED. IT WAS REPORTED THAT THE TALENT AORTO-UNI-ILIAC DEVICE WAS PLACED WITH A LEFT TO RIGHT FEM-FEM BYPASS. IN 2005, A PROXIMAL CUFF WAS PLACED FOR AN UNK REASON. THE PT HAS BEEN FOLLOWED BY ULTRASOUND, AND A TYPE 1 ENDOLEAK WAS IDENTIFIED; THEN, A CT SCAN SHOWED THAT THE DEVICE HAD MIGRATED INTO IN THE ANEURYSM SAC. AT THE TIME OF THE MIGRATION, THE NECK WAS FULL OF THROMBUS AND MEASURES 34MM IN DIAMETER, AND ESSENTIALLY THERE IS NO NECK LENGTH. THE AORTIC BIFURCATION IS VERY TORTUOUS AND THERE IS DISEASE PROGRESSION EXTENDING UPWARD TO THE VISCERAL VESSELS. A SURGICAL REPAIR IS PLANNED FOR A LATER DATE. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED, AND THE PT IS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TALENT ABDOMINAL STENT GRAFT SYSTEM | MIH | MEDTRONIC CARDIOVASCULAR SANTA ROSA | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Required Intervention |