TALENT ABDOMINAL STENT GRAFT SYSTEM - HYDRO
Report
- Report Number
- 2953200-2010-02566
- Event Type
- Injury
- Date Received
- December 16, 2010
- Date of Event
- November 17, 2010
- Report Date
- November 17, 2010
- Manufacturer
- MEDTRONIC CARDIOVASCULAR GALWAY
- Product Code
- MIH
- PMA / PMN Number
- P070027
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). EVALUATION: RESULTS: (ENDOLEAK), (SEVERELY ANGULATED AND REVERSE FUNNEL SHAPED AORTIC NECK), (STENT GRAFT WAS IMPLANTED IN A PT WITH A SEVERELY ANGULATED AND REVERSE FUNNEL SHAPED AORTIC NECK). CONCLUSION: (SEVERELY ANGULATED AND REVERSE FUNNEL SHAPED AORTIC NECK), (STENT GRAFT WAS IMPLANTED IN A PT WITH A SEVERELY ANGULATED AND REVERSE FUNNEL SHAPED AORTIC NECK).
A TALENT STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF A 5 CM ABDOMINAL AORTIC ANEURYSM APPROX 4 MONTHS AGO. THE AORTIC NECK WAS SHORT, REVERSE-CONICAL SHAPED AND RANGED FROM 20-28-32 MM IN DIAMETER OVER A LENGTH OF 15 MM, WITH SEVERE ANGULATION OF 70-80 DEGREES. IT WAS REPORTED THAT AT A ROUTINE F/U ON (B)(6) 2010, A SLIGHT PROXIMAL TYPE 1 ENDOLEAK WAS SEEN, LIKELY DUE TO THE SEVERE NECK ANATOMY. A 32 TALENT CUFF AND PALMAS STENT WERE PLACED TO SUCCESSFULLY RESOLVE THE ENDOLEAK. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED, AND THE PT IS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TALENT ABDOMINAL STENT GRAFT SYSTEM - HYDRO | MIH | MEDTRONIC CARDIOVASCULAR GALWAY | NA | V00490955 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR | Required Intervention |