FDA Adverse Event Injury Summary report: N

TALENT ABDOMINAL STENT GRAFT SYSTEM - HYDRO

MDR report key: 1933733 · Received December 16, 2010

Report

Report Number
2953200-2010-02566
Event Type
Injury
Date Received
December 16, 2010
Date of Event
November 17, 2010
Report Date
November 17, 2010
Manufacturer
MEDTRONIC CARDIOVASCULAR GALWAY
Product Code
MIH
PMA / PMN Number
P070027
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION: RESULTS: (ENDOLEAK), (SEVERELY ANGULATED AND REVERSE FUNNEL SHAPED AORTIC NECK), (STENT GRAFT WAS IMPLANTED IN A PT WITH A SEVERELY ANGULATED AND REVERSE FUNNEL SHAPED AORTIC NECK). CONCLUSION: (SEVERELY ANGULATED AND REVERSE FUNNEL SHAPED AORTIC NECK), (STENT GRAFT WAS IMPLANTED IN A PT WITH A SEVERELY ANGULATED AND REVERSE FUNNEL SHAPED AORTIC NECK).

Description of Event or Problem · 1

A TALENT STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF A 5 CM ABDOMINAL AORTIC ANEURYSM APPROX 4 MONTHS AGO. THE AORTIC NECK WAS SHORT, REVERSE-CONICAL SHAPED AND RANGED FROM 20-28-32 MM IN DIAMETER OVER A LENGTH OF 15 MM, WITH SEVERE ANGULATION OF 70-80 DEGREES. IT WAS REPORTED THAT AT A ROUTINE F/U ON (B)(6) 2010, A SLIGHT PROXIMAL TYPE 1 ENDOLEAK WAS SEEN, LIKELY DUE TO THE SEVERE NECK ANATOMY. A 32 TALENT CUFF AND PALMAS STENT WERE PLACED TO SUCCESSFULLY RESOLVE THE ENDOLEAK. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED, AND THE PT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TALENT ABDOMINAL STENT GRAFT SYSTEM - HYDRO MIH MEDTRONIC CARDIOVASCULAR GALWAY NA V00490955

Patients

Seq Age Sex Outcome Treatment
1 79 YR Required Intervention