FDA Adverse Event Malfunction Summary report: N

INFUSOMAT ®

MDR report key: 19335242 · Received May 16, 2024

Report

Report Number
9610825-2024-00370
Event Type
Malfunction
Date Received
May 16, 2024
Report Date
May 16, 2024
Manufacturer
B BRAUN MELSUNGEN AG
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN INTERNAL REPORT NUMBER (B)(4). NOTE: THIS REPORT IS BEING FILED FOR AN ITEM NUMBER THAT IS NOT SOLD IN THE UNITED STATES, HOWEVER SIMILAR ITEMS ARE SOLD IN THE UNITED STATES BY B. BRAUN MEDICAL, INC. PROBLEM: DELIVERY ACCURACY OUT OF TOLERANCE UNDERINFUSION 1. GENERAL INFORMATION: COMPLAINT: (B)(4). ---------------------------------------------------------------- 2. INFORMATION TO THE SAMPLE: 2.1 MODEL: INFUSOMAT SPACE 2.2 ARTICLE NUMBER: 8713050 2.3 SERIAL NUMBER/BATCH: (B)(6) 2.4 SOFTWARE VERSION: N030005 2.5 HOURS OF OPERATION: 8735 2.6 FURTHER INFORMATION: N/A ---------------------------------------------------------------- 3. INVESTIGATION RESULTS: 3.1 HISTORY INSPECTION: THE DEVICE HISTORY FILES WERE READ AND ANALYZED. THE DEVICE HISTORY FILES FROM (B)(6) 2024 WERE INVESTIGATED. BETWEEN 01:15AM AND 06:15AM COULD BE FOUND A INFUSION WITH RATE OF 70ML/H AND A VOLUME OF 1000ML. THE INFUSION STARTED AT 01:09AM AND WAS STOPPED AT 06:20AM WITHOUT ANY ABNORMALITIES. AT THIS TIME, 362,83ML WAS INFUSED. THE LINE WAS EXTRACTED. NO OTHER ABNORMALITIES WAS FOUND. 3.2 VISUAL INSPECTION: A VISUAL INSPECTION WAS PERFORMED. THE COVER CAPS ON THE SCREW PILLARS, AND THE PRODUCTION SEAL ON THE LOWER HOUSING WERE INTACT AND UNDAMAGED. THE DEVICE IS IN A CLEAN STATE AND NO VISIBLE DAMAGE ARE TO LOCATE. 3.3 FUNCTIONAL INSPECTION: A FUNCTIONAL TEST WAS PERFORMED. THE DEVICE PASSES THE SELF-TEST. A SPACE LINE WAS INSERTED, AND THE PUMP IDENTIFIED THE LINE, AND IT COULD BE SELECTED FROM THE MENU. IT WAS POSSIBLE TO PUT THE PUMP IN OPERATION. 3.4 PRESSURE INSPECTION: IN CHECKING THE DOWNSTREAM-SENSOR THE ELECTRONIC PRESSURE CUT-OFF AND THE MECHANICAL PRESSURE LIMITATION OF THE DEVICE WERE TESTED, ACCORDING TO THE REQUIREMENTS OF THE TECHNICAL SAFETY CHECK. THE DEVICE MATCHES THE REQUIRED VALUES AND STANDARDS. ALL MEASURED VALUES ARE WITHIN OUR SPECIFICATION. 3.5 FLOW RATE INSPECTION: A DELIVERY ACCURACY MEASUREMENT ACCORDING TO IEC 60601-2-24 WAS ARRANGED. HERE A NOMINAL FLOW RATE OF 100 ML/H WAS CHOSEN. THE ASSESSED AVERAGE DEVIATION "A" OF THE SECOND OPERATING HOUR WAS MEASURED AND RESULTED IN A VALUE OF +0,55%. ((ACCURACY OF SET DELIVERY RATE SHOULD BE: ± 5 % ACCORDING TO IEC/EN 60601-2-24) THE DEVICE MATCHES THE REQUIRED VALUES AND STANDARDS. ALL MEASURED VALUES ARE WITHIN OUR SPECIFICATION. 3.6 DISASSEMBLING: DURING THE INVESTIGATION NO FAULTS COULD BE DETECTED, TO INVESTIGATE THE INSIDE OF THE DEVICE, ONLY THE UPPER HOUSING WAS REMOVED. NO DAMAGE OR SOILING COULD BE FOUND. INFUSOMAT SPACE OPERATES WITHIN OUR SPECIFICATION. NO PRODUCT DEVIATION.

Description of Event or Problem · 0

ACCORDING TO THE CUSTOMER THE PUMP DELIVERED THE WRONG FLOW AND VOLUME. REPORTEDLY THE PUMP WAS SET AT 70 MILLILITERS AN HOUR (ML/H). DURING A CHECK IT WAS EXPECTED THAT 350 MILLILITERS (ML) WOULD HAVE BEEN DELIVERED HOWEVER, ONLY 100ML HAS GONE IN. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
618295 INFUSOMAT ® PUMP, INFUSION FRN B BRAUN MELSUNGEN AG 8713050

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown