FDA Adverse Event Malfunction Summary report: N

ALINITY C PROCESSING MODULE

MDR report key: 19335018 · Received May 16, 2024

Report

Report Number
3016438761-2024-00295
Event Type
Malfunction
Date Received
May 16, 2024
Date of Event
April 30, 2024
Report Date
June 10, 2024
Manufacturer
ABBOTT LABORATORIES
Product Code
JJE
UDI-DI
00380740137380
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE FIELD SERVICE REPRESENTATIVE (FSR) INSPECTED THE INSTRUMENT AND RESOLVED THE ISSUE BY REPLACING THE ICT MODULE. RETURN TESTING WAS NOT COMPLETED AS RETURNS WERE NOT AVAILABLE. THE INSTRUMENT SERVICE HISTORY REVIEW FOR (B)(6) REVEALED NO ADDITIONAL SERVICE TICKETS ASSOCIATED WITH DISCREPANT/ERRATIC RESULTS. THERE WERE NO ADDITIONAL SERVICE OR COMPLAINT ISSUES ON OR AROUND THE DATE THIS COMPLAINT WAS INITIATED THAT MAY HAVE CONTRIBUTED TO THIS ISSUE. A REVIEW OF TRACKING AND TRENDING FOR THE ALINITY C PROCESSING MODULE DID NOT IDENTIFY ANY TRENDS. A REVIEW OF TRACKING AND TRENDING FOR THE ICT MODULE DID NOT IDENTIFY ANY TRENDS. REVIEW OF THE MANUFACTURING DOCUMENTATION DID NOT IDENTIFY ANY NON-CONFORMANCES ASSOCIATED WITH THE COMPLAINT ISSUE. LABELING WAS REVIEWED AND FOUND TO ADEQUATELY ADDRESS THE ISSUE UNDER REVIEW. BASED ON THE INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY OF THE ALINITY C PROCESSING MODULE FOR SERIAL (B)(6) OR THE ICT MODULE WERE IDENTIFIED.

Additional Manufacturer Narrative · 0

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. ALL AVAILABLE PATIENT INFORMATION HAS BEEN INCLUDED. NO ADDITIONAL PATIENT DETAILS ARE AVAILABLE.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED FALSELY ELEVATED SODIUM RESULTS GENERATED ON THE ALINITY C PROCESSING MODULE. THE FOLLOWING DATA WAS PROVIDED (NORMAL REFERENCE RANGE 135-145): SID ORIGINAL RESULT (#6), REPEATED RESULT (#8) (B)(6) 161, 139 (B)(6) 161, 136 (B)(6) 163, 141 (B)(6) 165, 141 (B)(6) 166, 141 (B)(6) 170, 139 (B)(6) 173, 138. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED FALSELY ELEVATED SODIUM RESULTS GENERATED ON THE ALINITY C PROCESSING MODULE. THE FOLLOWING DATA WAS PROVIDED (NORMAL REFERENCE RANGE 135-145): SID ORIGINAL RESULT (#6) REPEATED RESULT (#8) (B)(6) 161 139 (B)(6) 161 136 (B)(6) 163 141 (B)(6) 165 141 (B)(6) 166 141 (B)(6) 170 139 (B)(6) 173 138. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
618232 ALINITY C PROCESSING MODULE ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE JJE ABBOTT LABORATORIES 00380740137380

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown ICT MODLE, 09D28-04, UNKNOWN| ICT MODLE, 09D28-04, UNKNOWN