AUTOMATED PD SET W/CASSETTE4 PRONG
Report
- Report Number
- 1423500-2010-07069
- Event Type
- Malfunction
- Date Received
- December 20, 2010
- Date of Event
- November 26, 2010
- Report Date
- November 26, 2010
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(4). PER THE CUSTOMER THE SAMPLE WAS DISCARDED AND THE LOT NUMBER WAS UNKNOWN, THEREFORE NO EVALUATION OR BATCH REVIEW COULD BE PERFORMED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF BAXTER'S INVESTIGATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
(B)(4). THIS COMPLAINT FOR THE SYSTEM ERROR (SE) 2240 (AIR IN LINE) WAS NOT CONFIRMED DUE TO A LACK OF SAMPLE. THE ROOT CAUSE OF THE COMPLAINT WAS NOT DETERMINED. THE LOT NUMBER IS UNKNOWN THEREFORE A BATCH REVIEW WAS NOT PERFORMED. SIMILAR REPORTS HAVE BEEN RECEIVED BY BAXTER FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
THE CUSTOMER CONTACTED BAXTER'S (B)(4) REGARDING SYSTEM ERROR 2240 ALARM THAT APPEARED ON THE HOMECHOICE (HC) DURING USE DURING THE LAST DWELL. THE BAXTER TECHNICAL SERVICE REPRESENTATIVE (TSR) ASSISTED THE HOME PATIENT (HP) WITH TROUBLESHOOTING, CLEARED THE ALARM, EXPLAINED THE ALARM AND ADVISED TO NOTIFY THE PERITONEAL DIALYSIS NURSE (B)(6). THE HP WOULD FINISH THAT NIGHT'S THERAPY MANUALLY. PROPER PROCEDURES PER THE USER MANUAL WERE REVIEWED WITH THE HP. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT. DURING FOLLOW-UP WITH THE HP, THE HP EXPLAINED THAT SHE WAS ASLEEP WHEN THE ALARM OCCURRED. SHE STATED THAT SHE DID NOT KNOW WHAT THE CAUSE OF IT WAS. THE HP HAD NOTIFIED THE NURSE. NO INFECTIONS/INJURIES HAD OCCURRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AUTOMATED PD SET W/CASSETTE4 PRONG | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 51 YR | HOME CHOICE |