LIGAMAX-5MM
Report
- Report Number
- 3005075853-2010-07268
- Event Type
- Malfunction
- Date Received
- December 20, 2010
- Report Date
- November 10, 2010
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K050344
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). JAW OR CAM INTERFACE IS DISENGAGED THE ANALYSIS RESULTS OF THE DEVICE FOUND THAT IT WAS RECEIVED WITH ONE JAW AND CAM RAMP DISENGAGED. THIS CONDITION WOULD NOT ALLOW THE JAWS TO COLLAPSE IN ORDER TO FORM THE CLIPS. THE DEVICE WAS CYCLED AND EJECTED THE REMAINING CLIPS DUE TO THE FOUND CONDITION. POSSIBLE CAUSES FOR THE FOUND CONDITION OF THE JAWS MAY BE INADVERTENT FORCE, TWISTING OR PRESSURE BEING PLACED ON THE DEVICE JAWS, USING THE JAWS OF THE DEVICE AS A DISSECTOR/RETRACTOR OR DAMAGE TO THE JAWS WHILE ENTERING THE TROCAR. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE,"THE DEVICE MISFIRED." THERE WAS NO PATIENT CONSEQUENCE REPORTED. THERE IS NO FURTHER INFORMATION AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIGAMAX-5MM | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |