FDA Adverse Event Malfunction Summary report: N

BD INSYTE AUTOGAURD

MDR report key: 19330836 · Received May 16, 2024

Report

Report Number
1710034-2024-00447
Event Type
Malfunction
Date Received
May 16, 2024
Date of Event
May 2, 2024
Report Date
June 6, 2024
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
Product Code
FOZ
UDI-DI
00382903825233
PMA / PMN Number
K201075
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.3. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED. B3. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD.

Additional Manufacturer Narrative · 0

INVESTIGATION RESULTS: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 382523 AND LOT NUMBER 4012523. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR QUALITY TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD INSYTE AUTOGAURD NEEDLE PIERCED THE CATHETER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WE HAVE HAD A RUN OF IV CATHETERS THAT ARE CAUSING ISSUES AT TIMES DEFECTIVE PLASTIC CATHETER-WHERE THE NEEDLE HAS GONE THROUGH THE PLASTIC SHEATH THAT COVERS THE NEEDLE. IV RESTARTS REQUIRED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
474405 BD INSYTE AUTOGAURD PERIPHERAL IV CATHETERS FOZ BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. 4012523 00382903825233

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown