FDA Adverse Event Malfunction Summary report: N

LIGAMAX-5MM

MDR report key: 1933027 · Received December 20, 2010

Report

Report Number
3005075853-2010-07259
Event Type
Malfunction
Date Received
December 20, 2010
Report Date
October 26, 2010
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K050344
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ADVANCER BYPASS. THE ANALYSIS RESULTS OF THE DEVICE FOUND THAT IT WAS RECEIVED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON FIRING, THE TIP OF THE ADVANCER SLID TOWARD TISSUE STOP DURING THE CONSECUTIVE FIRING SEQUENCES; PEAR SHAPE CLIPS WERE RELEASED. DURING THE NEXT FIRING SEQUENCE, THE CLIP WAS FORMED AS CONFORMING. THE DEVICE LOCKED OUT AS INTENDED BUT THE ORANGE INDICATOR WAS OVER TRAVELED. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE DEVICE MISFIRED. IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED. NO OTHER DETAILS OF THE EVENT ARE KNOWN. THERE WAS NO PATIENT IMPACT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIGAMAX-5MM CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA G4TU9P

Patients

Seq Age Sex Outcome Treatment
1