FDA Adverse Event Malfunction Summary report: N

ARCHITECT I2000SR ANALYZER

MDR report key: 1932992 · Received December 20, 2010

Report

Report Number
1628664-2010-00491
Event Type
Malfunction
Date Received
December 20, 2010
Date of Event
December 3, 2010
Report Date
December 6, 2010
Manufacturer
ABBOTT MANUFACTURING, INC.
Product Code
JJE
PMA / PMN Number
K983212
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). FURTHER INVESTIGATION OF THE CUSTOMER ISSUE INCLUDED A REVIEW OF THE COMPLAINT TEXT, A SEARCH FOR SIMILAR COMPLAINTS, AND A REVIEW OF LABELING. NO ERROR CODES WERE GENERATED WHEN THE FALSE (B)(6) HCV RESULT WAS GENERATED. TRACKING AND TRENDING DID NOT IDENTIFY ANY ADVERSE TREND IN CONJUNCTION WITH THE COMPLAINT ISSUE. LABELING WAS REVIEWED AND FOUND TO BE ADEQUATE. BASED ON THE DATA AVAILABLE AND THE RESULTS OF THIS EVALUATION NO PRODUCT DEFICIENCY WAS IDENTIFIED.

Additional Manufacturer Narrative · 1

(B)(4). A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE. AN INVESTIGATION IS IN PROCESS

Description of Event or Problem · 1

THE CUSTOMER STATED AN ARCHITECT (B)(4) ANALYZER GENERATED A (B)(6) RESULT FOR ONE PATIENT SAMPLE KNOWN TO BE A CARRIER OF (B)(6). THE ARCHITECT ANALYZER GENERATED AN INITIAL (B)(6) RESULT OF 0.03 S/CO AND A REPEAT RESULT OF 13.39 S/CO. THERE WAS NO ADVERSE IMPACT TO PATIENT MANAGEMENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ARCHITECT I2000SR ANALYZER AUTOMATED IMMUNOASSAY ANALYZER JJE ABBOTT MANUFACTURING, INC.

Patients

Seq Age Sex Outcome Treatment
1 ANTI-HCV REAGENT LN 6C37-20 LOT 88508HN00| ANTI-HCV REAGENT LN 6C37-20 LOT 88508HN00