FDA Adverse Event Malfunction Summary report: N

PUMP 1886 780G OUS BLE PUMP MG/DL

MDR report key: 19329152 · Received May 16, 2024

Report

Report Number
2032227-2024-171388
Event Type
Malfunction
Date Received
May 16, 2024
Date of Event
May 6, 2024
Report Date
October 21, 2024
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZP
UDI-DI
000076300063927301
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SA
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. SELECT PATIENT INFORMATION CANNOT BE PROVIDED DUE TO REGIONAL PRIVACY REGULATIONS. THE REPORTED DEVICE IS NOT MARKETED IN THE UNITED STATES, BUT IT IS A SAME/SIMILAR DEVICE TO ONE THAT IS MARKETED OUTSIDE THE UNITED STATES. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

INSERTED A TEST SC1-CAP INTO THE RETAINER RING, AND IT LOCKED IN PLACE PROPERLY. THE PUMP WAS RECEIVED WITHOUT A BATTERY CAP. THE PUMP PASSED SELF TEST; IT FAILED REWIND TEST RETURNING A PUMP ERROR 39. UNABLE TO PERFORM THE DISPLACEMENT, PRIME/SEATING, BASIC OCCLUSION, FORCE SENSOR, AND OCCLUSION TESTS DUE TO THE RECURRING PUMP ERROR 39S DURING MOTOR REWIND. THUMP SOFTWARE WAS UTILIZED AND UPLOADED TRACE/HISTORY FILES PROPERLY. THE CASE WAS CUT OPEN. AFTER VISUAL INSPECTION, DID NOT NOTE ANY SIGNS OF PREVIOUS MOISTURE PRESENCE OR CORROSION INSIDE THE BATTERY COMPARTMENT OR ON ANY OF THE BOARDS, ASSEMBLIES, AND THE MOTOR INSIDE THE PUMP. THE MOTOR WAS TESTED OUTSIDE THE DEVICE, AND IT FAILED DURING REWIND RETURNING A PUMP ERROR 39. A KNOWN GOOD MOTOR WAS PLUGGED INTO THE ORIGINAL BOARDS/CASE, AND THE MOTOR WAS ABLE TO REWIND COMPLETELY. DID NOT NOTE ANY CRACKS OR DISCONNECTS IN EITHER THE MOTOR OR FORCE SENSOR FLEX CABLES. IN SUMMARY, THE CUSTOMER¿S REPORT OF PUMP ERROR 39S WAS CONFIRMED DURING TESTING. THE PUMP ERROR 39S WERE ISOLATED TO THE MOTOR ASSEMBLY. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED PUMP ERROR 39. THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-1886. TROUBLESHOOTING WAS PERFORMED FOR PUMP ERROR 37-3941-4379-8082130. CUSTOMER WAS ABLE TO CLEAR THE ERROR BUT WAS UNABLE TO COMPLETE THE REWIND PROCESS. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. MMT-1886 WAS REQUESTED AND CUSTOMER RESPONSE WAS THE DEVICE WILL BE RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
924941 PUMP 1886 780G OUS BLE PUMP MG/DL AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL OZP MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1886 NG3524587H 000076300063927301

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown