FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 1932842 · Received December 20, 2010

Report

Report Number
2649622-2010-13236
Event Type
Malfunction
Date Received
December 20, 2010
Date of Event
September 22, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S017
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NM, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE PROXIMAL CONDUCTOR WAS DISTORTED DURING ANALYSIS, THERE WAS BLOOD/BODY FLUID ON THE DISTAL CONDUCTOR AND ON THE PROXIMAL CONDUCTOR, THE OUTER INSULATION WAS BREACHED CUT, THE HELIX WAS DISTORTED/BENT, AND THERE WAS BLOOD IN/ON THE HELIX MECHANISM; APPARENT EXPLANT DAMAGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD HAD HIGH IMPEDANCE VALUES IN DIFFERENT POSITIONS DURING IMPLANT ATTEMPT. THE LEAD WAS NOT USED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 86 YR Other