FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 19327973 · Received May 16, 2024

Report

Report Number
9610877-2024-53646
Event Type
Malfunction
Date Received
May 16, 2024
Date of Event
April 26, 2024
Report Date
May 16, 2024
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
EOB
PMA / PMN Number
K951196
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR. OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CFB (COHERENT FIBER BUNDLE) BROKEN. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE CFB (COHERENT FIBER BUNDLE). IN ADDITION, OUR TECHNICIAN CONFIRMED THAT THE U/D KNOB BROKEN, THE INSERTION FLEXIBLE TUBE BUCKLED, THE LIGHT GUIDE CABLE FOR PRONG BUCKLED, AND THE INSERTION FLEXIBLE TUBE LEAK; HOWEVER, THESE DEFECTS ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT ""HR-RPT-0586(IMAGE FAILURE)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(BROKEN).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
475197 PENTAX FIBER NASO PHARYNGO LARYNGOSCOPE EOB HOYA CORPORATION PENTAX TOKYO OFFICE FNL-10RP3

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown