Narratives
THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. SINCE THE DEVICE IS MORE THAN 15 YEARS OLD, THE DEVICE HISTORY RECORD WAS UNABLE TO BE REVIEWED. HOWEVER, OLYMPUS ONLY RELEASES PRODUCTS TO MARKET THAT MEET ALL MANUFACTURING SPECIFICATIONS AND FINAL PRODUCT RELEASE CRITERIA. BASED ON THE RESULTS OF THE INVESTIGATION, A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED. HOWEVER, IT IS LIKELY DUE TO THE FAILURE OF THE DIGITAL VISUAL BOARD THAT THE IMAGE IS NOT DISPLAYED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.
THE DEVICE EVALUATION IS ONGOING. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE VIDEO SYSTEM CENTER EXHIBITED INTERNAL FAILURE OF DV (DIGITAL VIDEO) BOARD CAUSING TERMINAL RATTLING AND NO DIGITAL OUTPUT. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.