FDA Adverse Event Malfunction Summary report: N

VISERA VIDEO SYSTEM CENTER

MDR report key: 19327669 · Received May 15, 2024

Report

Report Number
3002808148-2024-04489
Event Type
Malfunction
Date Received
May 15, 2024
Report Date
June 4, 2024
Manufacturer
SHIRAKAWA OLYMPUS CO., LTD.
Product Code
FET
UDI-DI
04953170141546
PMA / PMN Number
K051645
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. SINCE THE DEVICE IS MORE THAN 15 YEARS OLD, THE DEVICE HISTORY RECORD WAS UNABLE TO BE REVIEWED. HOWEVER, OLYMPUS ONLY RELEASES PRODUCTS TO MARKET THAT MEET ALL MANUFACTURING SPECIFICATIONS AND FINAL PRODUCT RELEASE CRITERIA. BASED ON THE RESULTS OF THE INVESTIGATION, A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED. HOWEVER, IT IS LIKELY DUE TO THE FAILURE OF THE DIGITAL VISUAL BOARD THAT THE IMAGE IS NOT DISPLAYED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Additional Manufacturer Narrative · 0

THE DEVICE EVALUATION IS ONGOING. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE VIDEO SYSTEM CENTER EXHIBITED INTERNAL FAILURE OF DV (DIGITAL VIDEO) BOARD CAUSING TERMINAL RATTLING AND NO DIGITAL OUTPUT. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
617932 VISERA VIDEO SYSTEM CENTER VIDEO SYSTEM CENTER FET SHIRAKAWA OLYMPUS CO., LTD. OTV-S7V 04953170141546

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown