INTEGRATED APD SET W/CASSETTE3-PRONG
Report
- Report Number
- 1423500-2010-07059
- Event Type
- Malfunction
- Date Received
- December 20, 2010
- Date of Event
- November 26, 2010
- Report Date
- November 26, 2010
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
(B)(4). THIS COMPLAINT FOR A SYSTEM ERROR 2240 (AIR IN SET) OCCURRED DURING INITIAL DRAIN WAS NOT CONFIRMED DUE TO A LACK OF SAMPLE; HOWEVER, PER THE COMPLAINT INFORMATION THE CAUSE OF THE SE 2240 IS DUE TO AIR BEING SUCKED INTO THE DISPOSABLE BECAUSE THE PATIENT EXTENSION LINE WAS CONNECTED AFTER PRIMING WAS COMPLETE. THE HOME PATIENT STATED THE PATIENT EXTENSION LINES WERE NOT CONNECTED PRIOR TO PRIMING. THIS REVIEW FINDS THE LABELING ADEQUATE FOR THE USE ERROR(S) IN THE COMPLAINT. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
(B)(4). THE SAMPLE WAS DISCARDED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON THE COMPLETION OF BAXTER'S INVESTIGATION.
A HOME PATIENT (HP) CONTACTED (B)(4) TO REPORT A SYSTEM ERROR (SE) 2240 ALARM THAT OCCURRED ON THE HOMECHOICE (HC) UNIT DURING INITIAL DRAIN. THE HP STATED THE PATIENT EXTENSION LINES WERE NOT CONNECTED PRIOR TO PRIMING. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ADVISED THE HP THAT AIR HAD ENTERED SETUP. THE TSR INSTRUCTED THE HP TO CYCLE POWER TO CLEAR THE ALARM AND FURTHER ADVISED TO START OVER WITH NEW SUPPLIES. THE TSR REVIEWED PROPER PROCEDURES PER THE USER MANUAL. THE TSR ADVISED THE HP TO CONNECT THE PATIENT LINE EXTENSION PRIOR TO PRIMING. THE HP CONFIRMED TO START OVER WITH NEW SUPPLIES. ON (B)(6) 2010 (B)(4) SPOKE WITH THE HP'S NURSE WHO STATED THAT THE HP DID NOT DEVELOP ANY ADVERSE EVENT OR REQUIRE ANY MEDICAL INTERVENTION. THE NURSE STATED THE HP WAS CURRENTLY PERFORMING THERAPY WITHOUT ANY COMPLICATIONS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INTEGRATED APD SET W/CASSETTE3-PRONG | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 86 YR | HOMECHOICE APD CYCLER |