FDA Adverse Event Malfunction Summary report: N

WAVESCAN WAVEFRONT SYSTEM

MDR report key: 1932728 · Received November 24, 2010

Report

Report Number
3006695864-2010-00133
Event Type
Malfunction
Date Received
November 24, 2010
Date of Event
July 30, 2010
Report Date
July 30, 2010
Manufacturer
AMO MANUFACTURING USA, LLC
Product Code
HKO
PMA / PMN Number
K000327
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE AMO FIELD SERVICE ENGINEER INSPECTED THE VIDEO BOARD FROM THE WAVESCAN COMPUTER AND FOUND THE CAPACITORS HAD BURNT AND WERE THE CAUSE OF THE BOARD FAILURE. NO FURTHER INFO IS AVAILABLE.

Description of Event or Problem · 1

AN AMO FIELD SERVICE TECH REPORTED THAT WHILE INSPECTING THE WAVESCAN AT THE CUSTOMER LOCATION FOR A REPORT OF NON OPERATION, HE NOTICED VISIBLE SMOKE MARKS ON THE BACK OF THE COMPUTER UNIT. INTERNALLY, HE FOUND THE VIDEO BOARD WAS BURNT AT THE CORNER AND BOTTOM AND THE MOTHERBOARD HAD EVIDENCE OF MELTING AT THE BUS SLOT. THE CUSTOMER INDICATED THAT THE SMOKE ALARM HAD BEEN TRIGGERED, HOWEVER, THEY DID NOT ASSOCIATE THIS WITH THE WAVESCAN. NO ONE AT THE CUSTOMER LOCATION REPORTED WITNESSING THE WAVESCAN SMOKING. THE FIELD SERVICE ENGINEER INDICATED THAT THERE WAS A SMELL OF BURNT ELECTRONICS IN THE ROOM WHERE THE EQUIPMENT WAS INSTALLED. AMO BELIEVES THAT THE WAVESCAN PCB COMPONENT FAILING RESULTED IN ENOUGH SMOKE BEING EMITTED TO HAVE TRIGGERED THE SMOKE ALARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 WAVESCAN WAVEFRONT SYSTEM HKO AMO MANUFACTURING USA, LLC WS1 NA

Patients

Seq Age Sex Outcome Treatment
1