FDA Adverse Event
Malfunction
Summary report: N
DASH PORT II
MDR report key: 1932688
·
Received December 14, 2010
Report
- Report Number
- MW5018603
- Event Type
- Malfunction
- Date Received
- December 14, 2010
- Report Date
- December 14, 2010
- Manufacturer
- GENERAL ELECTRIC
- Product Code
- MHX
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- OH, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Description of Event or Problem · 1
THE DOCKING STATION FOR THE DASH SERIES OF MONITORS FROM GE WILL INTERMITTENTLY DROP OFF THE PT CARE NETWORK. THIS PROBLEM LEAVES THE PT UNMONITORED AT A CENTRAL LOCATION. NO ALARMS REPORTING CAN OCCUR. GE'S FIX IS TO ROUTINELY CLEAN THE MONITOR CONTACTS THAT CONNECT TO THE DOCKING STATION. I FEEL THIS IS A POOR DESIGN, A POOR SOLUTION TO THE PROBLEM, AND THIS DESIGN IS A PT SAFETY ISSUE. GE ONLY SENT A LETTER TO THOSE THAT WERE COMPLAINING ABOUT THE PROBLEM. I FELT THE LETTER SHOULD HAVE ALERTED EVERYONE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DASH PORT II | DOCKING STATION | MHX | GENERAL ELECTRIC | DASH PORT II |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |