UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER
Report
- Report Number
- 2050012-2010-01690
- Event Type
- Malfunction
- Date Received
- December 20, 2010
- Date of Event
- November 19, 2010
- Report Date
- November 19, 2010
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
SAMPLES WERE RAN IN PLASMA TUBES FROM THE AUTOMATION LINE. QC HAD BEEN RUNNING WITHIN ACCEPTABLE ESTABLISHED SPECIFICATIONS PRIOR TO THE EVENT. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED AND FLUSHED THE GLUCOSE CUP AND ASSOCIATED TUBING LINES THE DAY BEFORE THE EVENT. THE FSE ALSO REPLACED THE REAGENT SYRINGE PUMP AND SAMPLE SYRINGE. ROOT CAUSE IS UNKNOWN TO DATE.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) REGARDING TWO (2) GLUCOSE (GLU) PATIENT SAMPLES DID NOT MATCH WHEN RUN IN DUPLICATE MODE THAT WAS GENERATED BY THE UNICEL DXC 800 PRO SYNCHRON CHEMISTRY ANALYZER. THE ORIGINAL RESULTS WERE ERRONEOUSLY LOW. THE RESULTS PROVIDED BY THE CUSTOMER ARE SHOWN. THE ERRONEOUS RESULTS WERE NOT REPORTED OUT OF THE LABORATORY. PATIENT TREATMENT WAS NOT AFFECTED IN THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | DXC 800 PRO | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |