FDA Adverse Event Malfunction Summary report: N

FG 6301 DUAL CHNL VOL INF PUMP2M8064 SERVICE

MDR report key: 1932571 · Received December 18, 2010

Report

Report Number
6000001-2010-05971
Event Type
Malfunction
Date Received
December 18, 2010
Date of Event
November 1, 2010
Report Date
November 24, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915523
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PR
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 1

(B)(4). A REQUEST FOR THE RETURN OF THE DEVICE HAS BEEN MADE. SHOULD THE DEVICE BE RECEIVED BY BAXTER FOR EVALUATION, A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF AN EVALUATION OR IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THIS DEVICE WAS RETURNED TO BAXTER FOR EVALUATION. A VISUAL INSPECTION AND FUNCTIONAL TESTS WERE PERFORMED. DEVICE EVALUATION CONFIRMED THE REPORTED CONDITION OF A FALSE OCCLUSION ALARM. THE ROOT CAUSE WAS DETERMINED TO BE A DEFECTIVE CENTRAL PROCESSING UNIT PRINTED CIRCUIT BOARD OCCLUSION CIRCUIT. NO REPAIRS WERE PERFORMED AS THE CUSTOMER REFUSED THE SERVICE ESTIMATE FOR THIS DEVICE. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. A FOLLOW UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A FLO-GARD INFUSION PUMP WHICH EXPERIENCED AN OCCLUSION ALARM UPON POWER UP. THIS CONDITION MAY BE A FALSE OCCLUSION ALARM. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.