FDA Adverse Event Malfunction Summary report: N

6461500 INFUSOR PUMP (1 LABEL)

MDR report key: 1932553 · Received December 18, 2010

Report

Report Number
6000001-2010-05966
Event Type
Malfunction
Date Received
December 18, 2010
Date of Event
November 1, 2010
Report Date
November 3, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K833577
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). IN THE INITIAL MEDWATCH REPORT IT WAS STATED "THERE IS ADVERSE EVENT, PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT." THIS STATEMENT IS INACCURATE. THERE IS NO ADVERSE EVENT, PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT.

Additional Manufacturer Narrative · 1

(B)(4). A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED WHEN THE EVALUATION RESULTS OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION WAS CONFIRMED AND DUPLICATED. THE ASSIGNABLE CAUSE COULD NOT BE DETERMINED AT THIS TIME. THE DEVICE WAS RETURNED TO THE CUSTOMER UNREPAIRED. ADDITIONAL: A SERVICE HISTORY REVIEW REVEALED THAT THIS DEVICE WAS NOT PREVIOUSLY SERVICED FOR THE REPORTED CONDITION. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED CONDITION. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE CONTACTED BAXTER TO REPORT AN INFUSOR PUMP WITH A CONDITION OF "ALARMS IF YOU PUT IN BOLUS MODE, BUT IT WILL RUN ON INFUSE". THE EVENT OCCURRED DURING PREVENTATIVE MAINTENANCE IN THE BIOMEDICAL SERVICE AREA. DURING THE PRODUCT EVALUATION AT BAXTER, IT WAS DISCOVERED WHEN A BOLUS IS STARTED, THE DEVICE RUNS FOR A FEW SECONDS AND THE GOES INTO A CONSTANT ALARM. THERE IS ADVERSE EVENT, PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT. THERE IS NO FURTHER COMPLAINT INFORMATION AVAILABLE.