COLLEAGUE SINGLE CHANNEL VOLUMETRIC PUMP CE FRENCH
Report
- Report Number
- 6000001-2010-05932
- Event Type
- Malfunction
- Date Received
- December 18, 2010
- Date of Event
- October 16, 2010
- Report Date
- October 26, 2010
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- Removal / Correction Number
- 6000001-12/1/08-002-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT CONTAIN A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US. A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED WHEN THE EVALUATION RESULTS OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION WAS CONFIRMED AND DUPLICATED. THE ASSIGNABLE CAUSE WAS DEPLETED MAIN BATTERIES. THE MAIN BATTERIES AND HARNESS HAVE BEEN REPLACED. ADDITIONAL: BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED CONDITION. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH MDQ-CAPA-(B)(4). THE USER INTERFACE MODULE MASTER SOFTWARE VERSION IS 5.08.92, CATEGORIZED AS REMEDIATED.
DURING AN ONSITE VISIT, A BAXTER FIELD SERVICE TECHNICIAN SERVICED A COLLEAGUE INFUSION PUMP FOR A DAMAGED BATTERY. THERE WAS NO ADVERSE EVENT, PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT. DURING BAXTER'S REVIEW OF THE EVENT HISTORY IT WAS DISCOVERED THAT A BATTERY DEPLETED SET ALARM OCCURRED DURING INFUSION WHICH CAUSED AN INTERRUPTION DURING DELIVERY. THERE IS NO FURTHER COMPLAINT INFORMATION AVAILABLE. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION IS CURRENTLY UNKNOWN.