FREESTYLE FREEDOM LITE
Report
- Report Number
- 2954323-2010-01694
- Event Type
- Malfunction
- Date Received
- December 17, 2010
- Date of Event
- November 29, 2010
- Report Date
- January 20, 2011
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE COMPLAINT WAS NOT CONFIRMED AND NO NEW ISSUES WERE OBSERVED. ALL TEST RESULTS WERE WITHIN RANGE SPECIFICATION. NO ERRORS WERE OBSERVED DURING CONTROL SOLUTION TESTING. THE READINGS THE CUSTOMER REPORTED WERE FOUND IN METER MEMORY. THIS IS A FINAL REPORT.
THE CUSTOMER'S PRODUCT HAS BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION IS OBTAINED. NOTE: THE METER IS DESIGNED TO DISPLAY READINGS OF 20 MG/DL TO 500 MG/DL. THIS METER DOES NOT SHOW A NUMERIC VALUE GREATER THAN 500 MG/DL. A BLOOD GLUCOSE READING ABOVE 500 MG/DL WILL READ AS A "HI" IN THE ADC METER.
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER WITHIN 10 MINUTES. RESULTS OF 501 MG/DL, 501 MG/DL, 156 MG/DL AND 142 MG/DL WERE PLOTTED ON THE PARKES ERROR GRID. THE RESULTS FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FREESTYLE FREEDOM LITE | BLOOD GLUCOSE MONITORING SYSTEM | NBW | 1014817 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|