FDA Adverse Event Malfunction Summary report: N

8MM MCS TIP COVER ACCESSORY

MDR report key: 1932278 · Received December 17, 2010

Report

Report Number
2955842-2010-00554
Event Type
Malfunction
Date Received
December 17, 2010
Date of Event
November 17, 2010
Report Date
November 19, 2010
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE MCS TIP COVER ACCESSORY HAS NOT BEEN RETURNED FOR EVALUATION. THE ROOT CAUSE OF THE CUSTOMER REPORTED FAILURE MODE CANNOT BE DETERMINED. IF THE TIP COVER IS RETURNED FOR EVALUATION, A FOLLOW-UP MDR WILL BE SUBMITTED. THE INSTRUMENT USED IN CONJUNCTION WITH THE TIP COVER WAS RETURNED AND EVALUATED. ENGINEERING EVALUATION FOUND THAT THE INSTRUMENT DOES NOT EXHIBIT EVIDENCE OF ARCING OR DAMAGE. THE TUBE EXTENSION AND GRIP SCISSOR TIPS DO NOT HAVE AN ARC TRACK. THE INSTRUMENT ALSO PASSED FUNCTIONAL TESTING.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI S HYSTERECTOMY PROCEDURE, ARCING WAS OBSERVED COMING FROM THE MONOPOLAR CURVED SCISSORS (MCS) INSTRUMENT WITH AN MCS TIP COVER ACCESSORY INSTALLED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 8MM MCS TIP COVER ACCESSORY ELECTROSURGICAL INSTRUMENT ACCESSORY NAY INTUITIVE SURGICAL,INC. 400180

Patients

Seq Age Sex Outcome Treatment
1 49 YR DA VINCI S SURG SYS, MCS INSTRUMENT, ACCSS, ESU