FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CXE1 PUMP UNREFURB

MDR report key: 1932265 · Received December 17, 2010

Report

Report Number
6000001-2010-05893
Event Type
Malfunction
Date Received
December 17, 2010
Date of Event
November 17, 2010
Report Date
November 17, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Removal / Correction Number
6000001-3/15/05-007-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THIS DEVICE WAS RETURNED TO BAXTER FOR EVALUATION. A VISUAL INSPECTION AND FUNCTIONAL TESTS WERE PERFORMED. DEVICE EVALUATION CONFIRMED THE REPORTED CONDITION OF A FALSE AIR IN LINE ALARM. THE ROOT CAUSE OF THE REPORTED CONDITION COULD NOT BE DETERMINED. NO REPAIRS WERE NECESSARY TO RESOLVE THE REPORTED CONDITION. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. A FOLLOW UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Description of Event or Problem · 1

DURING DEVICE EVALUATION, A BAXTER SERVICE TECHNICIAN REPORTED A COLLEAGUE INFUSION PUMP WHICH EXPERIENCED A CONSTANT, FALSE AIR IN LINE ALARM DURING THE ACCURACY TEST. THERE WAS NO PATIENT INVOLVED; THERE WERE NO REPORTS OF PATIENT INJURY OR MEDICAL INTERVENTION. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 6.13.90. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE CXE1 PUMP UNREFURB PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1