COLLEAGUE CXE1 PUMP UNREFURB
Report
- Report Number
- 6000001-2010-05893
- Event Type
- Malfunction
- Date Received
- December 17, 2010
- Date of Event
- November 17, 2010
- Report Date
- November 17, 2010
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Removal / Correction Number
- 6000001-3/15/05-007-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
(B)(4). DEVICE EVALUATION: THIS DEVICE WAS RETURNED TO BAXTER FOR EVALUATION. A VISUAL INSPECTION AND FUNCTIONAL TESTS WERE PERFORMED. DEVICE EVALUATION CONFIRMED THE REPORTED CONDITION OF A FALSE AIR IN LINE ALARM. THE ROOT CAUSE OF THE REPORTED CONDITION COULD NOT BE DETERMINED. NO REPAIRS WERE NECESSARY TO RESOLVE THE REPORTED CONDITION. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. A FOLLOW UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
(B)(4). BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
DURING DEVICE EVALUATION, A BAXTER SERVICE TECHNICIAN REPORTED A COLLEAGUE INFUSION PUMP WHICH EXPERIENCED A CONSTANT, FALSE AIR IN LINE ALARM DURING THE ACCURACY TEST. THERE WAS NO PATIENT INVOLVED; THERE WERE NO REPORTS OF PATIENT INJURY OR MEDICAL INTERVENTION. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 6.13.90. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE CXE1 PUMP UNREFURB | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |