FDA Adverse Event Death Summary report: N

CAPSUREFIX NOVUS

MDR report key: 1932009 · Received December 17, 2010

Report

Report Number
2649622-2010-15336
Event Type
Death
Date Received
December 17, 2010
Date of Event
January 23, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THE PATIENT DIED EIGHTEEN DAYS AFTER DEVICE REPLACEMENT. THE CAUSE OF DEATH HAS BEEN REQUESTED AND NOT RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 89 YR Death