FDA Adverse Event Malfunction Summary report: N

PCA3 MEDNET WIRELESS

MDR report key: 1931805 · Received November 22, 2010

Report

Report Number
2921482-2010-00921
Event Type
Malfunction
Date Received
November 22, 2010
Date of Event
January 1, 2010
Report Date
October 26, 2010
Manufacturer
HOSPIRA, INC.
Product Code
MEA
PMA / PMN Number
K042800
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THE DEVICE DISPLAYED E626 (AUDIBLE ALARM) WITHOUT SOUNDING AN AUDIBLE ALARM TONE. THE DEVICE WAS RETURNED TO THE BIOMEDICAL ENGINEERING DEPT WITH AN UNSIGNED NOTE THAT STATED, "NO ALARM, E626 DISPLAYED ON SCREEN." NO TRACKING INFO WAS PROVIDED; THEREFORE, SPECIFIC PT INFO, DEVICE PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. DURING TESTING AT THE USER FACILITY, THE DEVICE DID NOT SOUND AN AUDIBLE ALARM TONE DURING AN ALARM CONDITION. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. MULTIPLE UNSUCCESSFUL ATTEMPTS HAVE BEEN MADE TO OBTAIN ADDITIONAL INFO INCLUDING EVENT DETAILS AND EVENT DATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PCA3 MEDNET WIRELESS 80MEA MEA HOSPIRA, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK