FDA Adverse Event Malfunction Summary report: N

ENDOTAK ENDURANCE EZ

MDR report key: 19315285 · Received May 14, 2024

Report

Report Number
2124215-2024-29376
Event Type
Malfunction
Date Received
May 14, 2024
Date of Event
April 19, 2024
Report Date
May 14, 2024
Manufacturer
BOSTON SCIENTIFIC CORPORATION
Product Code
LWS
PMA / PMN Number
P910073/S026
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

KADOSAKA, TAKEHIDE, ET AL. (2024). NORMALIZATION OF INCREASING SHOCKING COIL IMPEDANCE WITH FULL OUTPUT SYNCHRONIZED SHOCK. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY MANUSCRIPT ID JCE-24-0337.

Description of Event or Problem · 0

IT WAS REPORTED THAT TECHNICAL SERVICES (TS) WAS CONTACTED BY THE HEALTH CARE PROFESSIONAL (HCP) TO REVIEW A JOURNAL ARTICLE. IN THE ARTICLE, THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH, OUT-OF-RANGE SHOCK IMPEDANCE MEASUREMENTS OF GREATER THAN 200 OHMS. CHEST RADIOGRAPHY SHOWED NO MACROSCOPIC APPEARANCES OF A LEAD INSULATION BREACHES. AFTER A DELIVERED SHOCK, THE IMPEDANCE MEASURED 65 OHMS AND REMAINED THERE FOR FIVE YEARS. TS AND THE HCP DISCUSSED IT WAS LIKELY THE SINGLE OUT-OF-RANGE IMPEDANCE MEASUREMENT WAS THE RESULT OF ELECTROMAGNETIC INTERFERENCE (EMI). THE RV LEAD REMAINS IN SERVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
476064 ENDOTAK ENDURANCE EZ IMPLANTABLE LEAD LWS BOSTON SCIENTIFIC CORPORATION 0154

Patients

Seq Age Sex Outcome Treatment
1 60 YR Male