SYRINGE 0.3ML 31GA 8MM TW 10BAG 500 CA
Report
- Report Number
- 3024508819-2024-00055
- Event Type
- Malfunction
- Date Received
- May 14, 2024
- Report Date
- August 8, 2024
- Manufacturer
- EMBECTA MEDICAL I LLC - HOLDREGE, NE
- Product Code
- FMF
- PMA / PMN Number
- UNKNOWN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
ADDITIONAL INFORMATION WAS ADDED TO: B4, G6, H2, H3, H11. CORRECTION TO: H6 (MEDICAL DEVICE CODE, INVESTIGATION FINDINGS, AND INVESTIGATION CONCLUSIONS) INVESTIGATION SUMMARY: NO PHYSICAL SAMPLES WERE RECEIVED HOWEVER THE INVESTIGATION WAS PERFORMED BASED ON THE PHOTO(S) PROVIDED. THIS IS THE 1ST COMPLAINT FOR THE REPORTED LOT NUMBER. A REVIEW OF THE MANUFACTURING RECORDS WAS PERFORMED, AND NO NON-CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS TYPE OF EVENT FOR THIS LOT. EMBECTA WAS ABLE TO DUPLICATE OR CONFIRM THE INDICATED ISSUE AND BASED ON TREND ANALYSIS NO FURTHER ACTION IS REQUIRED AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. IF SAMPLES ARE RECEIVED IN THE FUTURE THE COMPLAINT WILL BE REOPENED FOR FURTHER INVESTIGATION. BASED ON THE ABOVE, NO ADDITIONAL INVESTIGATION AND NO CORRECTIVE/PREVENTATIVE ACTION (CAPA) OR SITUATION ANALYSIS (SA) IS REQUIRED AT THIS TIME.
ACCOUNT - DERMETICS ONTARIO CANADA MATERIAL - SYRINGE 0.3ML 31GA 8MM TW 10BAG 500 CA MATERIAL# - 320440 LOT# - 4002004 COMPLAINT - WE HAVE NOT NOTICED THIS FLUID COMING FROM NEW SYRINGES BEFORE. THE FLUID FEELS SLIPPERY AND VISCOUS, NOT LIKE WATER, BUT LIKE AN OILY TEXTURE. ALSO, TO REITERATE WHAT I SAID ON THE PHONE, WE ARE AN EXTREMELY HIGH-VOLUME CLINIC, AND NEED TO HAVE THIS MATTER INVESTIGATED AND ADDRESSED AS PRIORITY. YOU HELP IN ESCALATING THIS MATTER AND FACILITATING A PROMPT RESPONSE IS VERY APPRECIATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1102795 | SYRINGE 0.3ML 31GA 8MM TW 10BAG 500 CA | SYRINGE, PISTON | FMF | EMBECTA MEDICAL I LLC - HOLDREGE, NE | 320440 | 4002004 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |