ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
Report
- Report Number
- 2531779-2010-02791
- Event Type
- Malfunction
- Date Received
- December 8, 2010
- Report Date
- November 11, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: PUMP BOOTS UP TO 'VERIFY' SCREEN WITH FUNCTIONAL AUDITORY AND VIBRATORY ALARMS. NO VISUAL/ DAMAGE OBSERVED TO KEYPAD. "CONTRAST BUTTON UNRESPONSIVE AND DOES NOT SPRING BACK WHEN PRESSED. 'UP/OK' BUTTONS WERE RESPONSIVE WHILE THE 'DOWN' BUTTON WAS INTERMITTENTLY RESPONSIVE; 'UP/DOWN/OK' BUTTONS SPRING BACK NORMALLY. DURING A VISUAL INSPECTION, THERE WAS CONTAMINATION OBSERVED UNDER 'DOWN/CONTRAST' BUTTONS AND NO CONTAMINATION OBSERVED UNDER 'UP/OK' BUTTONS.
THE LAY-USER/PT ALLEGED THAT THE BUTTONS ON THE KEYPAD DO NOT RESPOND. THERE WAS PRODUCT MISUSE. THE KEYPAD DID NOT APPEAR TO BE PEELED OR TORN. FURTHERMORE, THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ONE TOUCH PING |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |